Measurement
See the Optical Tip Table in Appendix for a complete list of optical tips.

Measurement Procedure

Attach a measurement optical tip, position the probe for maximum
accuracy and freeze the image. Using the
left soft key,
select Measurement. Choose the measurement tip and the type of
measurement. Position the cursor and press
Freeze/Enter to save
the cursor placement. Repeat with remaining cursor(s).

Capturing Measurement Images for Re-Measurement

Stereo or Shadow measurement images can be saved allowing the user
to perform the actual measurements at a later time on the XL Vu, Rhythm
Review station, or on a PC using the Inspection Manager software
Measurement Procedure
Attach a measurement optical tip, position the probe for maximum
accuracy and freeze the image to be saved for re-measurement. Press
the joystick's Main Menu button. Select 'Save As" and press
select
'Measurement' and press
. Complete other optional fields and
press
. When complete, use the
soft key, select DONE.

Accuracy Index

The accuracy index is a number that indicates the amount of
magnification during shadow or stereo measurements. The higher
the accuracy index, the greater the magnification. The greater the
magnification, the more accurately you can place the measurement
cursors allowing for more accurate results.
®
98