Chapter 3 Features and Functions
Calibration Overview
To Secure Against Calibration You can secure the function
generator against accidental calibration either from the front panel or
over the remote interface. The function generator is secured when
shipped from the factory and the security code is set to "HP033120".
Be sure to read the security code rules on page 118 before attempting to
secure the function generator.
• Front-Panel Operation:
1: UNSECURED
(CAL MENU)
If the function generator is unsecured, you will see the above
command when you go into the
generator, go to the "
PARAMETER
command, enter the security code, and then press Enter.
∧000000:CODE
When you go to the "
COMMANDS
you will notice that the function generator is secured. Notice also that
the "
" command is now hidden and you cannot perform a
CALIBRATE
calibration.
1: SECURED
• Remote Interface Operation:
CALibration:SECure:STATe {OFF|ON},<
To secure the function generator, send the above command with the
same code as used to unsecure calibration. For example,
"CAL:SEC:STAT ON,HP033120"
120
. To secure the function
CAL MENU
" level of the
UNSECURED
" level in the
CAL MENU
>
code
again,