Field Mountable Videographic Recorder
SM500F
2.2.1 Panel-Mounting – Fig. 2.4
1
2
Fig. 2.4 Installing the Instrument – Panel-Mounting
Note. If removal of knockouts is required, refer to Section 2.3.1 on page 13.
Referring to Fig. 2.4, secure the instrument in a panel as follows:
a
Cut the correct sized hole in the panel – see Fig. 2.3 on page 8.
b
Insert the instrument into the panel cut-out.
c
Screw one clamping screw into the left-hand bracket until 10 to 15mm of the thread protrudes from
the other side of the bracket and position one clamp over the end of the thread.
d
Holding the assembly together, position the bracket into the left-hand recess on the rear of the
instrument case and secure with the bracket securing screw. Ensure that the plastic washer remains
in the position fitted.
Repeat instructions
e
Tighten the clamping screws evenly and securely by hand.
Note. This is critical in order to ensure proper compression of the panel seal and achieve the
IP66/NEMA 4X hosedown rating.
IM/SM500F Issue 10.1
Left-hand panel clamp assembly only shown.
Right-hand assembly omitted for clarity
3
4
and
for the right-hand panel clamp assembly.
Bracket
Securing
Clamp
Bracket
Screw
4
3
2 Installation
Clamping Screw
5
9

Hide quick links: