7.2

ReadyToProcess WAVE 25 rocker

Rocking
Error symptom
Rocker does not initial-
ize properly.
When starting the run,
the rocker moves too
slowly during the first
cycle.
Rocker is not rocking.
Rocker stops rocking.
Incorrect rocking angle.
Rocking speed varies.
ReadyToProcess WAVE 25 Operating Instructions 29009597 AD
Possible cause
Faulty safety fuses.
It takes a while for the
rocker to initialize after
turning on the power.
Rocker is in an error
state.
Tray is in tilt position.
Rocker is mechanically
restricted from moving.
DO control using speed
is enabled. If the DO
deviates from the
setpoint, the rocking
speed changes.

7.2 ReadyToProcess WAVE 25 rocker

Corrective action
Try to restart the system a few times.
If the system still does not initialize properly, shut
it down and contact GE service personnel.
None. This is normal.
Check the current alarms.
Reset power to the unit.
If you still get motor alarms, contact GE service
personnel.
Make sure that the tray is in normal position.
If this does not help, contact GE service person-
nel.
Locate and remove the restricting object.
This is normal when DO control is enabled. If not
required, disable DO control from the process pic-
ture.
7 Troubleshooting
157