5-9. Dynamic Range Test Setup - HP 1317A Operating And Service Manual

Table of Contents

Advertisement

Performance Check
Model1317A
Model 1317A
SQUARE WAVE
GENERATOR
IRIGGER
OUTPUT
50n
OUTPUT
y
INPUT
(IN REAR)
MEASURE VERTICAL HEIGHT
OF ROUNDING
X
INPUT
(IN REAR)
MAIN SWEEP
OUTPUT
(IN REAR)
MONITOR OSCILLOSCOPE
0'
@
0
IS>
0
@
Ioii
/
I
VERTICAL
INPUT
/
C)
"
10
-
.:..18
I
I
1317 A -L-013
Figure
5-9. Dynamic Range Test Setup
e.
Set oscillator frequency
to
I MHz
an
d
ampli-
tude
t
o display full screen raster on CRT.
f. Set square wave generator for
100-kHz,
50-volt output at A5TPl.
5-10
g. Set monitor oscilloscope
to
observe leading
edge of displayed waveform with display exactly
filling vertical graticule.
h. Observe risetime between vertical 10% and
90% points. Risetime shall be less than 20 ns.