Process Validation - Miele PG 8581 Operating Instructions Manual

Dental washer-disinfector
Table of Contents

Advertisement

Maintenance measures
Test point for
measuring
sensors
104

Process validation

The standard of cleaning and intermediate-level thermal disinfection
in the disinfection programs must be confirmed by the user as a
routine matter.
Safety checks and performance validation must be carried out in
accordance with the internationally recognized standard EN ISO
15883. In some countries additional national regulations, guidelines
and recommendations also apply.
The sensor test point for validation is located at the front right on the
top of the machine under the lid or the countertop. To reach the
access point, the lid of the machine must be removed or the machine
must be pulled out from under the countertop.
Open the door.
Unscrew the retaining screws.
Then remove the safety screws on the back of the machine from
the lid and lift the lid to remove it.
Or
Pull the machine out by approx. 6" (15 cm) from under the
countertop.