Download Print this page

Sanyo MLS-3751L Service Manual page 31

High-pressure steam sterllizer labo autoclave

Advertisement

(2) Adjustment of pressure sensors
Pressure correction of the pressure sensor can be performed manually.
Pressure correction is performed at 2 points, namely, in the status with the lid being opened and at 135 DC.
For prevention of heating without water, check the heating water amount.
I I
Operation
I I
I I
Description
II
I I
Display
I I
I
Enter into the test mode .
..
Fix the 2 shafts found at the back
side
of
the cover by tape, etc.
and set
it
to ON status.
t
Close the lid once and then open
it.
..
Press the START button with
the cover being opened.
The correction
of
pressure is
displayed only while the
reservation
button is oressed.
When the
cycle
selection
button
is pressed while pressing
it,
the display changes in the
order
of
-Lateral sensor-Object
temp. sensor-Bottom sensor-
Press. sensor. Select the
pressure correction display.
Press the
~or
"a"
button
while pressing the reservation
buttu
I
and adjust
the pressure
so that the pressure display valu
becomes 'P. 0'.
When the START
button
is
pressed while pressing the
reservatm
buttu
i,
the
corrected value is written in.
Press the
STOP
butIpn and
remove the tape from the shaft.
I
For adjusting the pressure when
the pressure is 0 (cover 'Open').
Selection of the pressure
correction display
(See the right figure.)
Pressure adjustment
(Correction)
Pressure adjustment
(Writing-in)
- 30-
Normal display
Release.
Raserv.
Press.
button
Example
of
]
nannal display
Cycle selection
button
~
Lateral thermistor correction
display
( Cycle selection
button
1
I
Object temp. correction display
[ Cycle selection
J
button
Bottom thermistor correction
display
Example of pressure
Cycle selection
J
correction display
button
' -
e.1
rn
( j
\
]
/
\
Pressure display value
Corrected value

Advertisement

loading

This manual is also suitable for:

Mls-3751Mls-3781Mls-3781l