Features <MDF-U333> <MDF-U333> <MDF-U333> <MDF-U333> ■ HFC refrigerant Refrigerant circuit contains HFC refrigerant that effect to environment little. Latest cooling system reduces negative factor to global environment. With ambient temperature at 35℃, while temperature of inside chamber maintains -30℃. ■ PCM structure PCM structure reduces amount of organic solvent.
Page 5
Features <MDF-U537,537D> <MDF-U537,537D> <MDF-U537,537D> <MDF-U537,537D> ■ HFC refrigerant Refrigerant circuit contains HFC refrigerant that effect to environment little. Latest cooling system reduces negative factor to global environment. With ambient temperature at 35℃, while temperature of inside chamber maintains -30℃. ■ PCM structure PCM structure reduces amount of organic solvent.
Page 6
Specification MDF-U333 Items Specification Model description Biomedical freezer Model Number MDF-U333 W 614 × D 733 × H 1620 mm Outer dimension (Max) W 490 × D 485 × H 1290 mm Inner dimension Effective chamber capacity 274 lit. Exterior...
Page 7
MDF-U333 Items Specification Microprocessor controlled system with non-volatilized memory. Temperature controller Temp. set range : -18℃~-40℃ Temperature sensor Thermistor LED digital display (green colored) by 1℃. Temperature display Range : -50℃~50℃ Changeable SV±5℃~15℃. (Initial SV±10℃) High/Low ALARM lamp brinks and buzzer sounds intermittently Temperature (about 15 min.
Page 8
Specification MDF-U537 Items Specification Model description Biomedical freezer Model Number MDF-U537 W 804 × D 772 × H 1802 mm Outer dimension (Max) W 658 × D 607 × H 1272 mm Inner dimension Effective chamber capacity 482 lit. Exterior Zinc galvanized steel w/powder polyester coating (PCM) Interior Polystyrene resin...
Page 9
MDF-U537 Items Specification Microprocessor controlled system with non-volatilized memory. Temperature controller Temp. set range : -18℃~-40℃ Temperature sensor Thermistor 2P LED digital display (green colored) by 1℃. Temperature display Range : -50℃~50℃ Changeable SV±5℃~15℃. (Initial SV±10℃) High/Low ALARM lamp brinks and buzzer sounds intermittently Temperature (about 15 min.
Specification MDF-U537D Items Specification Model description Biomedical freezer Model Number MDF-U537D W 804 × D 772 × H 1802 mm Outer dimension (Max) Upper chamber W 658 × D 607 × H 589 mm Inner dimension Lower chamber W 658 × D 607 × H 603 mm Effective chamber capacity 452 lit.
MDF-U537D Items Specification Microprocessor controlled system with non-volatilized memory. Temperature controller Temp. set range : -18℃~-40℃ Temperature sensor Thermistor 2P (upper and lower compartment) LED digital display (green colored) by 1℃. Range : -50℃~50℃ Temperature display Selectable to display upper or lower compartment by switch. Changeable SV±5℃~15℃.
Specifications of sensor The following show temperature and resistance characteristics on each thermistor sensor (type 502). Temperature Resistance Temperature Resistance Temperature Resistance (℃) (kΩ ) (℃) (kΩ ) (℃) (kΩ ) -50 -7 154.50 17.92 8.17 -45 -6 116.50 17.16 7.85 -40 -5...
Page 28
Control specification Control specification Control specification Control specification < < < < MDF-U333・ ・ ・ ・ 537・ ・ ・ ・ 537D> > > > 1. Key and switch Key and switch Key and switch Key and switch 1. 1. 1....
Page 29
In PV display keep pressing DEF key over 5 seconds, defrosting will be started. Model code 001 (MDF-U333) --- Force to turn compressor OFF. Model code 002 (MDF-U537) --- Force to turn compressor & solenoid valve OFF. (Use 1 solenoid valve) Model code 003 (MDF-U537D) --- Single compartment →...
Page 30
4. 4. 4. 4. Self diagnosis function E01 : Temperature sensor (upper compartment temperature sensor for MDF-U537D) is open circuit E02 : Temperature sensor (upper compartment temperature sensor for MDF-U537D) is short circuit E03 : Lower compartment temperature sensor is open circuit (for MDF-U537D only) E04 :...
Page 31
5. 5. 5. 5. Key lock function In PV display、Keep pushing key over 5 seconds to enter key lock mode displaying with “L0”. 1 ・・・・・key lock ON Push key to 0 ・・・・・key lock OFF then press SET key to memorize set value then return to PV display. Note)...
Page 32
F07 : To adjust zero for temperature sensor (Upper compartment sensor for MDF-U537D) F08 : To adjust zero for temperature sensor (Not available for MDF-U333) (Solenoid valve control sensor for MDF-U537) (Lower compartment sensor for MDF-U537D) F09 : Not available F10 :...
Page 33
F01 : To set temperature of high temp. alarm performance in the range of +5℃~+15℃. (initial value : +10℃) F02 : To set temperature of low temp. alarm performance in the range of -5℃~-15℃. (initial value : -10℃) F03 : To set temperature of high temp. alarm performance in the range of +5℃~+15℃. (initial value : +10℃) Return to PV display except for lower compartment of MDF-U537D.
Page 34
F08 :To adjust 0 of temperature sensor (setting range: -9.9℃~+9.9℃) In F17: It is functioned in model code 002(MDF-U537), 003(MDF-U537D). Return to PV display in model code 001(MDF-U333) Setting method: In ”F08” displayed, press SET key to display ”00.0”(initial value).
Page 35
F17 :Model code setting or checking 001 :MDF-U333 002 :MDF-U537 002 :MDF-U537D Followings are initial setting values Internal temp. SV : -30℃ (for U537D : both compartments) Key lock mode : 0 (OFF) High temperature alarm : SV+10℃(for U537D : both compartments) Low temperature alarm :...
Page 36
The function is to correct differential value between temperature of temperature sensor and of internal 1/2H (centre of chamber). Initial offset 001 (MDF-U333): Shift –1.5℃from temperature of temperature control sensor (PV) 002 (MDF-U537): Shift +1.0℃from temperature of temperature control sensor (PV) (For upper, lower compartment) 003 (MDF-U537D): Shift +0.0℃from temperature of temperature control sensor (PV)
Test data ・MDF-U333 Pull down ・MDF-U333 Pull up ・MDF-U333 Cooling unit pressure ・MDF-U333 Power consumption ・MDF-U333 Temperature uniformity and fluctuation ・MDF-U537 Pull down ・MDF-U537 Pull up ・MDF-U537 Cooling unit pressure ・MDF-U537 Power consumption ・MDF-U537D Pull down ・MDF-U537D Pull down ・MDF-U537D Pull up ・MDF-U537D Cooling unit pressure ・MDF-U537D Power consumption ・MDF-U537/D Temperature uniformity and fluctuation...
MDF-U333 Temperature uniformity and fluctuation (Test condition) Mains : 1φ100V50/60Hz : 35℃ : -25℃・-20℃・-30℃ No load Measure with type T sensor Measuring point:Center point of each container 1 2 3 4 5 (1).SV=-25℃ Measuring point Temperature(℃) Fluctuation(℃) -21.8 -23.2 3( 1/2 center air) ±...
MDF-U537Cooling unit pressure Pd(60Hz) Ps(50Hz) Pd(60Hz) Ps(60Hz) 2000 AT 35℃,no load 1500 1000 Time scale [H] MDF-U537 Power consumption W(50Hz) W(60Hz) AT 35℃,no load Time scale [H] − 44 −...
MDF-U537D Pull down AT35℃,50Hz,no load Top chamber Bottom chamber Time scale[H] MDF-U537D Pull down AT35℃,60Hz,no load Top chamber Bottom chamber Time scale[H] − 45 −...
MDF-U537 / D temperature uniformity and fuluctuation Mains : 100V50Hz : 35℃ :-20℃・-25℃・-30℃ No load Mesure with type T sensor Measureing point : Center point of each container /chamber MDF-U537 MDF-U537D 1 1 4 2 2 8 3 3 9 4 5...
Page 53
Parts layout MDF-U537 MDF-U537 MDF-U537 MDF-U537 Electrical parts on top Top view of product Access port for recorder wires Inside of control panel Connecter for buzzer Connecter for display P.C.B. Back view of product Door latch Remote alarm terminal Breaker switch Inner parts of door latch ・Temperature control sensor (MDF-U537)