Safety and Care

ACUSON Freestyle Rollstand Configuration

WARNING: When operating and transporting the ultrasound system using the
Rollstand, insure that the rollstand is configured according to the specifications provided
in this manual. Failure to do so can cause the system to tip over and create a risk for
injury to the user or patient and damage to the system.
4
6
Rollstand Configuration
1.
Quick Release Adaptor Plate: the top of the
48 in/122 cm. Do not place anything other than the Freestyle, probe batteries, and probes
on this mount.
2.
Handle: the top of the Handle should be placed no higher than 40 in/101 cm from the
floor.
3.
Keyboard Shelf: should be placed no higher than 33 in/84 cm from the floor.
4.
Rear Storage Basket: the top of the basket should be placed no higher than 30 in/76 cm
from the floor. Weight limit 4 lbs/1.8 kg.
5.
Front Storage Basket: top of basket should be placed no higher than 24 in/61 cm from the
floor. Weight limit 4 lbs/1.8 kg.
6.
Cable Wrap: should be placed no higher than 22 in/56 cm.
7.
Video Printer Shelf: the shelf should be placed no higher than 12 in/30 cm from the floor.
Weight limit 6 lbs/2.7 kg. Secure printer with straps.
2-44
should be placed no higher than
plate
Freestyle
ACUSON
1
2
3
5
7
Freestyle User Manual
ACUSON