Siemens SITRANS MAG 6000 I Operating Instructions Manual

Electromagnetic flowmeters
Hide thumbs Also See for SITRANS MAG 6000 I:
Table of Contents

Advertisement

Quick Links

SITRANS F
Electromagnetic flowmeters
SITRANS MAG 6000 I/6000 I Ex de
Operating Instructions
7ME693 (MAG 6000 I)
08/2018
A5E02083319-AH
Introduction
Safety notes
Description
Installing/Mounting
Connecting
Commissioning
Functions
Alarm, error, and system
messages
Service and maintenance
Troubleshooting/FAQs
Technical data
Spare parts/Accessories
Menu diagrams
Factory settings
Control drawing
1
2
3
4
5
6
7
8
9
10
11
12
A
B
C

Advertisement

Table of Contents
loading

Summary of Contents for Siemens SITRANS MAG 6000 I

  • Page 1 Introduction Safety notes Description SITRANS F Installing/Mounting Electromagnetic flowmeters SITRANS MAG 6000 I/6000 I Ex de Connecting Commissioning Operating Instructions Functions Alarm, error, and system messages Service and maintenance Troubleshooting/FAQs Technical data Spare parts/Accessories Menu diagrams Factory settings 7ME693 (MAG 6000 I)
  • Page 2 Note the following: WARNING Siemens products may only be used for the applications described in the catalog and in the relevant technical documentation. If products and components from other manufacturers are used, these must be recommended or approved by Siemens. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems.
  • Page 3: Table Of Contents

    Pipe or wall mounting with assembly bracket................27 Connecting..............................29 Safety precautions.........................30 Electrical connection......................31 Connection of sensor and transmitter in remote version............34 Wiring in hazardous area.......................37 Connection of bus communication add-on modules..............38 Reassembling the device.......................38 SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 4 Technical data............................65 11.1 Technical specifications......................65 11.2 Dimensions and weight......................67 11.3 Accuracy..........................68 11.4 Output characteristics......................69 11.5 Cable data..........................71 11.6 Cable requirements........................72 Spare parts/Accessories..........................73 12.1 Ordering of spare parts......................73 12.2 Accessories..........................73 SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 5 50 Hz Dimension dependent Qmax..................91 60 Hz Dimension dependent Qmax..................92 50 Hz Dimension dependent volume/pulse and batch............94 60 Hz Dimension dependent volume/pulse and batch............95 Control drawing............................97 Control drawing........................97 Index.................................103 SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 6 Table of contents SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 7: Introduction

    The following table lists the most important changes made to the documentation in the current and previous editions. Edition Remarks HW version SW version 08/2018 4.08 X06 ● Updated Safety Notes 04/2016 4.08 X06 ● Various technical specifications updated ● Control drawing updated SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 8: Device Identification

    2. Make sure the scope of delivery, and the information on the namelate corresponds to the ordering information SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 9: Further Information

    Product information on the Internet The Operating Instructions are available on the documentation disk shipped with the device, and on the Internet on the Siemens homepage, where further information on the range of SITRANS F flowmeters may also be found: Product information on the internet (http://www.siemens.com/flowdocumentation)
  • Page 10 Introduction 1.5 Further Information SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 11: Safety Notes

    ● Altitude up to 2000 m ● Maximum relative humidity 80% for temperatures up to 31°C (88 °F) decreasing linearly up to 50% relative humidity from 40 °C (104 °F) SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 12: Installation In Hazardous Area

    Material compatibility Siemens Flow Instruments can provide assistance with the selection of wetted sensor parts. However, the full responsibility for the selection rests with the customer and Siemens Flow Instruments can take no responsibility for any failure due to material incompatibility.
  • Page 13: Atex 2 Gd Approval

    These markings are for the Remote version only; the marking applicable to the Compact versions depend upon their construction. See the latest issues of certificates Sira 07ATEX3181X, Sira 07ATEX1182X and Sira 03ATEX3339X. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 14 Sensor electrode input (Terminals 0, 81, 82, 83, 84 - "ia circuits") 30 V DC 30 V DC 6.1 mA 6.1 mA 560 nF 66 nF 0.96 H 45.5 mW 45.5 mW SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 15 0 nF 36 µH 18.4 mH 4.6 mH 0 µH 36 µH 0 µH 0 µH 0.7 W 0.66 W 0.61 W 1.2 W 1.2 W 5.32 W 1.2 W SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 16 ● sensor and transmitter are connected to the potential equalizing conductor. ● when protective earth (PE) is connected, no potential difference between the protective earth (PE) and the potential equalization (PA) can exist, even during a fault condition. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 17: Fm/Csa Class I, Zone 1 Approval

    0 nF 36 µH 18.4 mH 4.6 mH 0 µH 36 µH 0 µH 0 µH 0.7 W 0.66 W 0.61 W 1.2 W 1.2 W 5.32 W 1.2 W SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 18: Fm Class I, Ii, Iii, Div 1 Approval

    3. Conduit seal is required within 460 mm (18") from MAG 6000 I in hazardous areas (Class I, Div. 1). The control drawings are found on the CD-ROM shipped with the device and on the Siemens homepage at http://www.siemens.com/flowdocumentation. See also...
  • Page 19: Description

    The SITRANS F M platform enables fitting of add-on bus modules without loss of analog, pulse and relay outputs, and all modules can be fitted as true Plug & Play. Standard transmitter: ● HART ● PROFIBUS PA and DP SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 20: Design

    ● Low flow cut-off ● Analog, pulse/frequence and relay outputs ● Optional: additional digital communication modules ● Error and error log system ● Uni-/bidirectional flow ● Limit switches ● Integrated batch control SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 21: Features

    The output module converts flow data to analog, digital and relay outputs. The outputs are galvanically isolated and can be individually set to suit a particular application. Theory of operation The flow measuring principle is based on Faraday’s law of electromagnetic induction. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 22 23 bit signal resolution. This has eliminated the need for range switching. The dynamic range of the transmitter is therefore unsurpassed with a turn down ratio of minimum 3000:1. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 23: Installing/Mounting

    As the transmitter housing and the display can be oriented in all directions, reading and operating the flowmeter is possible under almost any installation conditions. 360° To ensure optimum flow measurement, attention should be paid to the following: SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 24: Compact Installation

    The SENSORPROM is factory-mounted in the transmitter. ® Figure 4-2 Compact installation Note To avoid too high tension on small sizes MAG 1100 sensors, transmitter must be supported. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 25: Remote Installation

    Cable length and type (as described in Cable requirements (Page 72)) must be used. For installation conditions for sensors, see respective sensor operating instructions. 1. Remove sensor terminal box lid. 2. Remove SENSORPROM unit from sensor terminal box. ® SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 26 Electrode cable and coil cable must be kept separate to prevent interference. 6. Tighten cable glands and lid screws well to obtain optimum sealing. WARNING Mount terminal box lid before power up. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 27: Wall Mounting Using Standard Mounting Plate

    Pipe or wall mounting with assembly bracket Optional pipe or wall mounting assembly bracket Note The assembly bracket is not part of the standard delivery and must be ordered separately. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 28 2. Fasten the transmitter and assembly bracket to the wall Note The fastening brackets and nuts are not needed for wall mounting. Wall mounting with assembly bracket. Dimensions in mm (inch). SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 29: Connecting

    Cable loops in the terminal box must be avoided. ● To guarantee the IP 67/NEMA 4 degree of protection, use cables with external diameters matching the used cable glands. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 30: Safety Precautions

    35 A. WARNING Danger of electric shock Do not power up when lid is open. Do not open lid while power is on, unless operating in an Ex ia environment. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 31: Electrical Connection

    Connecting 5.2 Electrical connection Electrical connection Ω Ω Ω Ω Electrical connection label The electrical connection label is placed in the lid of the terminal box. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 32 Exception are sensor connections that are connected to ground. WARNING Grounding The mains protective earth wire must be connected to the PE terminal in accordance with the diagram (Protection class I). SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 33 Non-Ex versions can be connected as either active or passive. Note The factory default setting for the current output is marked by an "X" and can be found on a label on the back of the lid. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 34: Connection Of Sensor And Transmitter In Remote Version

    ® WARNING Danger of electric shock Do not power up when lid is open. Do not open lid while power is on, unless operating in an Ex ia environment. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 35 ® 2. Fit coil and electrode cables through cable glands in sensor terminal box. 3. Connect signal and coil cables to transmitter as follows: SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 36 Avoid exposing transmitter to direct sunlight Exposing the transmitter to direct sunlight may increase the operating temperature above its specified limit, and decrease display visibility. In such case, protect the transmitter from direct sunlight. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 37: Wiring In Hazardous Area

    = Σ R = Σ R < < Loop max Loop max - Loop - Loop Current output active with and without HART Barrier / Isolator < 650Ω SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 38: Connection Of Bus Communication Add-On Modules

    91-97. For more information Refer to the relevant BUS communication Quick Start or Operating Instructions available at the SITRANS F literature CD or on the internet, at: www.siemens.com/flowdocumentation (www.siemens.com/flowdocumentation) Reassembling the device Ensure that the two screws (20 TORX) are tightened properly for correct assembly and proper ground connection.
  • Page 39 Connecting 5.6 Reassembling the device Ensure sufficient insulation by covering power supply terminals with plastic cover. Fasten the plastic cover with two screws. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 40 Connecting 5.6 Reassembling the device SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 41: Commissioning

    Local user interface With the capacitive touch keypad the device can be operated without opening the lid. An LED light gives feedback on accepted command. Figure 6-1 Local user interface SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 42 FORWARD KEY This key is used to step forward through the menus. It is the only key normally used by the operator. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 43: Menu Structure

    1. Press top up key for 2 sec. 2. Enter password. 3. Use forward key or backward key to reach password menu. 4. Press lock/unlock key to unlock password. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 44: Changing Basic Settings

    Setting of velocity unit per time unit Error level Selecting error level at which flowmeter will detect an error. Note Totalizer 2 is not visible when batch is selected as digital output. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 45 6. Press lock/unlock key to unlock settings. 7. Use select key and change key to change settings. 8. Press lock/unlock key to confirm new settings. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 46 − − − − − − − Decimal point can be positioned and units set individually for flow rate in totalizer 1 and totalizer SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 47: Changing Operator Menu Setup

    4. Use change key to select requested reading. 5. Press lock/unlock key to confirm selection. 6. Use select key to move cursor to line 3. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 48: Changing Language

    5. Use change key to select desired language. 6. Press lock/unlock key to confirm new language. 7. Press top up key two times to exit setup mode. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 49: Functions

    3.5 mA High 22.6 mA For setting of error level, see Error level menu diagram (Page 79). If current output is not used, it must be set to "Off". SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 50: External Input

    By applying 11 … 30 V DC to terminals 77 and 78, it is possible to perform: ● Batch control (start, stop, hold/continue) ● Reset totalizer ● Force/freeze output ● Q 2 (night) See external input menu diagram (Page 82). SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 51: Sensor Characteristics

    It is possible to check whether the outputs are functioning. Error pending and status log lists are also accessible from this menu and the operating time (in days) can be read. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 52 Functions 7.6 Service mode The forced control is stopped and all previous settings are reinitialized the moment the service mode is left by pressing top up key SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 53: Alarm, Error, And System Messages

    The acceptance level for "error pending" can be individually configured to a particular application. The acceptance level is set in the basic settings menu (Page 44). SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 54 The alarm field, error output and error pending always operate together. Operator menu Error pending and status log are as default enabled ( ) in the operator menu. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 55: List Of Error Numbers

    Switch off and on A parameter is out of range. The parameter could not Active Active be replaced by its default value. The error will disap‐ pear at the next power-on SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 56 Active ® unit anymore F70 - Coil current Check cables Coil excitation has failed Active Active F71 - Internal error Replace transmitter Internal conversion error in ASIC Active Active SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 57: Service And Maintenance

    ● Seal integrity of the process connections, cable entries, and cover screws ● Reliability of power supply, lightning protection, and grounds NOTICE Repair and service must be carried out by Siemens authorized personnel only. Note Siemens defines flow sensors as non-repairable products.
  • Page 58: Return Procedures

    ● Information about field service, repairs, spare parts and lots more under Services. Additional Support If you have additional questions about the device, please contact your local Siemens representative and offices at: Local contact person (http://www.automation.siemens.com/aspa_app/contactmenu.aspx? ci=yes&regid=DEF&lang=en)
  • Page 59 You can find the forms on the Internet and on the CD delivered with the device. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 60 Service and maintenance 9.4 Return procedures SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 61: Troubleshooting/Faqs

    Ensure sufficient potential equalization tween medium and sensor 4. Air bubbles in medium Ensure medium does not contain air bubbles 5. High concentration of parti‐ Increase time constant cles or fibres SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 62: Transmitter Check List

    Display light on? Yes ⇒ 1 No ⇒ 2 Flashing error triangles? Yes ⇒ Check error table No ⇒ 1.2 Output and display readings OK? Yes ⇒ 1.2.1 No ⇒ 1.2.2 SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 63 No ⇒ 2.1.2. 2.1.1 Display defective Replace display 2.1.2 Transmitter defective Replace transmitter Note Sensor check list Check lists for sensors are included in the respective sensor operating instructions. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 64 Troubleshooting/FAQs 10.2 Transmitter check list SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 65: Technical Data

    Exceptions are sensor connections that are connected to ground. Max. measuring error MAG 6000 I/MAG 6000 I Ex de ±0.2% ± 1 mm/s (incl. sensor and zero point) SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 66 Ex d e [ia] ia IIC T6 Gb Ex tD A21 IP67 T85°C Ta -25°C to +60°C FM Class I, II, III, Div 1 FM Class I, Zone 1/21 CSA Class I, Zone 1/21 SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 67: Dimensions And Weight

    Dimensions and weight Transmitter IP67/NEMA 4X compact die cast aluminum Figure 11-1 Transmitter dimensions shown in mm (inch) Weight: MAG 6000 I/MAG 6000 I Ex de: 6 kg (13.5 lbs) SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 68: Accuracy

    As pulse output ± (0.1% of actual flow + 0.05% FSO) Effect of ambient temperature Display/frequency/pulse output < ± 0.003% / °C act. Current output < ± 0.005% / °C act. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 69: Output Characteristics

    Additions in the event of deviations from reference conditions 11.4 Output characteristics Current output Output characteristics Bidirectional mode Unidirectional mode 4 to 20 mA Frequency output Bidirectional mode Unidirectional mode Pulse output Bidirectional mode Unidirectional mode SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 70 Error relay output Bidirectional mode Unidirectional mode No error Error Limit switch (can be used as direction switch) 1 set point 2 set points Passive Digital output Passive Digital output SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 71: Cable Data

    Electrode cable, double shielded (for empty pipe detection or low conductivity fluids) Cable kit with standard coil cable and electrode cable double shielded (also available as low noise cable for MAG 1100 sensor) SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 72: Cable Requirements

    For Ex zone 1 installations only certified cable glands with protection type "e" can be used for the power supply and the coil cable. The cable glands must be approved for the actual temperature and cable dimension. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 73: Spare Parts/Accessories

    6000 I / MAG 6000 I Ex de HART FDK-085U0321 Modbus RTU/RS485 FDK-085U0234 PROFIBUS PA Profile 3 FDK-085U0236 PROFIBUS DP Profile 3 FDK-085U0237 Devicenet FDK-085U0229 FOUNDATION Fieldbus H1 A5E02054250 SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 74: Spare Parts

    Display unit FDK-085U3122 Electronics cover with Ex glass plate A5E02593565 Cover for connection board incl. gas‐ 7ME5933-0AC02 ket (for remote version) Cover for mains supply/communica‐ 7ME5933-0AC03 tion including gasket SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 75: Menu Diagrams

    Menu diagrams Transmitter menu overview The menu diagrams shown on the following pages apply to MAG 5000/6000 as well as MAG 6000 I. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 76 Menu diagrams A.1 Transmitter menu overview Not available in MAG 5000 Add-on module Factory-set password: 1000 Not available when batch Only available when batch SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 77: Basic Settings

    PDM or ordered by using the Z-option Y20 in the ordering system. The customer unit in the time unit list is displayed as "?". Via PDM or Y20 option this could e.g. be chosen as year "y". SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 78: Current Output

    Do not combine differentiation and low alarm level together with passive output. The output will try to pull down the level to 1.3 mA at fatal errors which is not possible for passive output. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 79: Digital Output - Pulse

    Menu diagrams A.6 Error level Digital output - pulse Digital output - frequency Error level SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 80: Error Number

    Menu diagrams A.8 Direction/limit Error number Direction/limit SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 81: Batch

    If digital output is set up for pulse, frequency or batch, then the relay output will be turned off if it has been set up for batch. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 82: External Input

    (The display is locked during this time). The time cycle can be set at 1 to 240 hours. If the cycle is set at for example three hours, the transmitter will be active every three hours. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 83: Sensor Characteristics

    In this case the power must be switched off/on to suppress the P40 error on the output. The frequency can be set to 44 Hz in the MAG 6000 SV transmitter only. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 84: Reset Mode

    Menu diagrams A.12 Reset mode A.12 Reset mode SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 85: Service Mode

    9 is the best signal measured. If digital output is set to pulse (standard). If digital output is set to frequency. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 86: Operator Menu Setup

    2 and flow velocity is chosen in line 3, the text "Flow velocity " is shown in line2 and the measured flow velocity is shown in line 3. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 87: Product Identity

    Menu diagrams A.15 Product identity A.15 Product identity SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 88: Add-On Communication Module

    Menu diagrams A.16 Add-on communication module A.16 Add-on communication module Example: HART Note Burst mode is not available with HART communication SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 89: Factory Settings

    0 to 9.9% Empty pipe On, Off Velocity unit m, mm, cm, dm, ft , in per s, min, h, d, ? (customer unit) Error level Warning Fatal, permanent, warning SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 90 0.85 to 2.00 Sensor charac‐ teristics teristics (Page 83) Language Language English English, German, French, Danish, Swedish, Fin‐ Changing lan‐ nish, Spanish, Russian, Italian, Portuguese, Pol‐ guage (Page 48) SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 91: 50 Hz Dimension Dependent Qmax

    15904.31 800 (32) 7000 452.3894 18095,57 452.3894 18095.57 900 (36) 9000 572.5553 22902,21 572.5553 22902.21 1000 (40) 12000 706.8584 28274.33 706.8584 28274.33 1050 (42) 12000 706.8584 28274.33 706.8584 28274.33 SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 92: 60 Hz Dimension Dependent Qmax

    1244.881 US GPM 125 (5) 44.05714 1762.285 48.62819 1945.127 US GPM 150 (6) 1101 69.32434 2772.973 70.02459 2800.984 US GPM 200 (8) 1762 109.7986 4391.941 124.48819 4979.525 US GPM SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 93 3000 (120) 308201 28009.84 1120393 US GPM * Factory setting sets Qmax to a metric unit (see previous table). The values here are converted to rounded off US GPM. SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 94: 50 Hz Dimension Dependent Volume/Pulse And Batch

    1500 (60) 1.472622 l 38603.89 m 1600 (66) 1.675517 l 43922.64 m 1800 (72) 2.120576 l 55589.6 m 2000 (78) 2.617994 l 68629.13 m 2200 (90) 3.167773 l 83041.25 m SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 95: 60 Hz Dimension Dependent Volume/Pulse And Batch

    300 (12) 0.01528871 400784.1 0.01556102 407922.7 350 (14) 0.02118028 555228.2 0.02118028 555228.2 400 (16) 0.02766404 725196 0.02766404 725196 450 (18) 0.0350123 917826.2 0.0350123 917826.2 500 (20) 0.04322506 1133118 0.04322506 1133118 SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 96 11603130 1800 (72) 0.5601967 14685210 2000 (78) 0.6916009 18129900 2200 (90) 0.836837 21937170 2400 (96) 0.995906 26107050 2600 (102) 1.168806 30639530 2800 (114) 1.355538 35534600 3000 (120) 1.556102 40792270 SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 97: Control Drawing

    Control drawing Control drawing SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 98 Control drawing C.1 Control drawing SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 99 Control drawing C.1 Control drawing SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 100 Control drawing C.1 Control drawing SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 101 Control drawing C.1 Control drawing SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 102 Control drawing C.1 Control drawing SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 103: Index

    Document History, 7 Nameplate, 9 Electrical connection, 38 Basic requirements, 34 Cable specifications, 29 In hazardous area, 37 Plug & Play, 21 Safety, 29 Pressure Error handling, 53 Safety instructions, 23 SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...
  • Page 104 Instrument safety standards, 11 Sensor installation, 23 Sensorprom, 21 Service, 57, 58 Settings, 49 Support, 58 System components, 19 Theory of operation, 21 Troubleshooting, 61 Wiring, (See Electrical connection) SITRANS MAG 6000 I/6000 I Ex de Operating Instructions, 08/2018, A5E02083319-AH...

Table of Contents