Phase Angle Rotation Correction Display And Operating Procedures - Toshiba LQ500 Operation Manual

Density meter
Hide thumbs Also See for LQ500:
Table of Contents

Advertisement

5.2.13

Phase angle rotation correction display and operating procedures

5:SET PARAMETERS
6:ZERO CALIBRATION
7:SPAN CALIBRATION
8:ANGLE ROTATION
Move the cursor to "8" with [→] key, and press [SET] key
to select [8:ANGLE ROTATION]
UH:UPPER ANGLE
DATA : 260
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
SH:LOWER ANGLE
DATA :100
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
N:ROTATION
DATA: 0
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
UH:UPPER ANGLE
DATA :260
[SET] CHANGE
[ESC] RETURN
[ESC]
(Previous
Menu)
[SET](Set)
UH:UPPER ANGLE
RANGE :240-360 °
[SET](Fix)
DATA : 260
[SET] SET,[ESC] CANCEL
[ESC](Cancel)
Before shipping, this value is set to 260 degrees. Changing
the setting may result in damaging the otherwise normal
operation. Therefore, normally, do not change the setting.
[SET](Set)
SH:LOWER ANGLE
RANGE :0-120
[SET](Fix)
DATA : 100
[SET] SET,[ESC] CANCEL
[ESC](Cancel)
Before shipping, this value is set to 100 degrees. Changing
the setting may result in damaging the otherwise normal
operation. Therefore, normally, do not change the setting.
[SET](Set)
N:ROTATION
RANGE :−10 - 10
[SET](Fix)
DATA : 00
[SET] SET,[ESC] CANCEL
[ESC](Cancel)
Normally, the angle rotation N is automatically computed;
therefore, it is unnecessary to change the setting through
manual input. However, there are some cases of high density
measurements which require settings through manual input.
 47 