Calibrating The Printer - Honeywell PX940 User Manual

High performance industrial printer
Hide thumbs Also See for PX940:
Table of Contents

Advertisement

If your verifier is not calibrated, follow these steps:
Note: Do not power off the printer during verifier calibration.

Calibrating the Printer

1.
2.
3.
4.
5.
Note: You should see the tear bar through the diamond cut on the calibration card black
line.
6.
7.
8.
Note: If Calibration fails, see
PX940 User Guide
ALIGN BLACK LINE WITH TEAR BAR
HPN: 50144664-001
PASTE LABEL AREA
From the front touch panel, go to: Wizards > Calibration > Verifier and follow
the instructions.
Open the verifier arm and print head arm.
Remove any media and ribbon from the printer media path.
Clean the printhead, platen roller, and verifier glass with a soft, lint-free cloth
soaked in isopropyl alcohol.
Place the calibration card on the printer media path, according of the
instructions written on the calibration card.
The three chevrons on the card indicating the orientation. The "FRONT" text
on the card placed outside of the verifier
Align the black line with the printer tear bar. Use the diamond cut out on the
calibration card black line to align to the tear bar.
Align the card evenly toward the printer spine and resting against the media
guide.
Close the verifier arm and then the printhead.
Press NEXT on the touch screen.
The card will move forward as the calibration process is running.
Once completed, the front panel will show if the calibration pass or fail.
Verifier Error Alerts and Solutions,
beginning on page 65.