Cannot Record For 24 Hours - GE SEER Light Service Manual

Ambulatory recorder/controller
Table of Contents

Advertisement

Cannot Record for 24 Hours

2-14
SEER Light/SEER Light Extend Recorder: Troubleshooting
Cannot record for
24 hours.
Are there many
artifacts in the ECG
recording?
No
Were the batteries
removed before recording
for 24 hours?
No
Were alkaline
batteries used?
Yes
Are the
battery electrodes dirty
or improperly
soldered?
No
Call GE Service.
SEER Light Ambulatory Recorder/Controller
2019818-008
Yes
Refer to flow chart on
artifact later in this chapter.
Yes
Don't remove the batteries
before recording for 24 hours.
No
Use alkaline
batteries.
Yes
Clean the battery electrodes
or improve soldering.
Revision B