Change the Resolution
To change the resolution or increase the response speed,
follow the procedure below.
1. Remove the cover (two M3 × 6 screws).
2. Reset the dip switches on the printed circuit board (Fig.
3-1. S101) by referring to Tables 3-1 and 3-2.
3. Reinstall the cover and fasten with screws.
4. Install the unit on the machinery and confirm
operations (counter status, etc.)
20
OFF
ON
Fig. 3-1
• Resolution switch setting: (Table 3-1.)
Switch setting
µ
Resolution (
m)
S101-1
100
OFF
50
20
ON
10
• Responce speed switch setting: (Table 3-2.)
Switch setting
µ
TW (
sec)
Output cycle (kpps)
S101-3
OFF
2.5 (100)
ON
5 (50)
OFF
10 (25)
ON
20 (12.5)
• Maximum response speed (m/min): (Table 3-3.)
µ
Tw (
s)
2.5
µ
Resolution (
m)
100
300
50
300
20
300
10
192
S101-2
OFF
ON
OFF
ON
S101-4
OFF
ON
20
5
10
300
300
240
300
240
120
192
96
48
96
48
24