GE eBike II Series Service Manual page 34

Table of Contents

Advertisement

Insert the jumper adapter into the corresponding
socket, to close all necessary connections.
Close the housing and fix to the ergometer chassis
again.
The COM module is plugged onto the 9 pole DSUB
connector (PORT 1) and screwed to the ergometer chassis.
The analog adapter has to be activated as follows:
In Service Menu go to System Configuration and set
Analog Interface to I.
Then power OFF/ON the eBike.
Go to Settings, select ECG Type, enter password 003
and select Analog / Digital
Insert the appropriate digital and analog connecting cables
and fix the cables to the chassis using a strain relief..
- 34 -
bKe-pcb
coM
coM
eBike II, eBike II L, eBike II EL
with JuMper adapter inStalled
9
dSub
Module Mounted on
pole
Module connectorS
2018112-069 Rev. F