Attach Canister Brackets To Head End Of The Secondary Base; Figure 6-120: Canister Bracket Assembly (With And Without Canister) - GE Discovery LS System Service Manual

Table of Contents

Advertisement

GE
HEALTHCARE
D
2307224-100, R
IRECTION
EVISION
6.2.7.7

Attach Canister Brackets to Head End of the Secondary Base

1.) Locate and remove the Right Canister Bracket Assembly, P/N 2332626-2, and the Left
Canister Bracket Assembly, P/N 2332626, from the upgrade kit. (The canister brackets are
mirror images.)
2.) Refer to
the gantry.
3.) Use a 5 mm Allen wrench to attach the Canister Bracket Assemblies to the front end of the
table with four M10 x 20 flat-head beveled screws (P/N 2103580P22).
6
Figure
6-120. Position the L-shaped canister bracket so the free end points away from

Figure 6-120: Canister Bracket Assembly (with and without Canister)

Chapter 6 - Table
D
ISCOVERY
Bracket assembly ships with
bottom bolt threaded through hole.
Hang the roller canister off this
bolt to support its weight while
fastening it into place.
Roller Cover Flaps
LS S
S
M
YSTEM
ERVICE
ANUAL
Page 195

Advertisement

Table of Contents
loading

Table of Contents