Technical Specifications; Measurements; Environmental - GE Intellix BMT 330 User Manual

Hide thumbs Also See for Intellix BMT 330:
Table of Contents

Advertisement

Technical Specifications

Measurements

Feature
Maximum Bushing Temperature at
sensor
Relative Phase Angle accuracy
Maximum number of PD measured
Maximum PD measurable

Environmental

Feature
Cabinet Operating Temperature
Range
When equipped with multi-
master comms option
Operating Humidity Range
Maximum Operating Altitude
Storage Temperature Range
Minimum Ambient Temperature for
Installation and Service Activities
Cabinet Protection Level
Bushing Sensor Protection Level
Intellix™ BMT 330 Weight
Bushing Sensor Weight
Bushing Sensor Operating
Temperature
Page 8
Intellix
Value
90 deg C (194 deg F)
0.01 deg of angle
200 PD per cycle (50 to 60Hz)
CAT III. 5Vac rms, 200mA; on each phase
Value
-40 to +55 deg C (-22 to +131 deg F)
-40 to +55 deg C (-22 to +131 deg F)
5 to 95% RH, non-condensing
2000m (6500 ft) above sea level
-40 to +85 deg C (-40 to +185 deg F)
-20 deg C (-4 deg F)
IP55
IP66
20 kg (44lb)
350 gm (0.8 lb), typical
-40 to +90 deg C (-40 to +194 deg F) at
Bushing Tapping Point
BMT330UM
v1.0
BMT 330 User Manual
20-Jun-16