Adjustment and test procedure Manual
4. Adjustment of DC voltage, AC voltage and DC current.
Specification
Item
DC
Voltage
AC
Voltage
DC
Current
5.Test
5-1 No Input on DC Voltage Mode
n
Specification : ±00.1
n
Set Rotary S/W to DC Voltage mode and check DM-311 to meet the above specification.
5-2 Short-circuit on DC Voltage Mode
n
Specification : ±00.1
n
Set Rotary S/W to DC Voltage mode and Short-circuit between COM terminal and V terminal
n
Check DM-311 to meet the above specification.
5-3 DC Voltage Measurement Test
Set Rotary S/W to DC Voltage Mode and Apply below voltage to terminal COM and terminal V to each
range with a calibrator
Version 1.0
LG Precision Co., Ltd.
1. Set Rotary S/W to DC Voltage Mode
189.8mV ~
2. Apply DC 190mV to Terminal COM and Terminal V of DM-311 with
190.2mV
a calibrator
3. Adjust VR1 to be displayed 190.0mV on LCD display
1. Set Rotary S/W to AC Voltage Mode
189.5mV ~
2. Apply AC 190mV/60Hz to Terminal COM and Terminal V of DM-311 with
190.5mV
a calibrator
3. Adjust VR2 to be displayed 190.0mV on LCD display
1. Set Rotary S/W to DC Current 10A Mode
9.95A ~
2. Apply DC 10A current to terminal COM and terminal 10A of DM-311
10.05 A
with a calibrator
3. Adjust Mn Wire(R39) with ripper to be displayed 10.00A on LCD display.
Customer
Range
Specification
DC 200mV
DC 2V
±(0.5% + 1dgt)
DC 20V
DC 200V
DC 1000V
Procedure
Factory
Specification
±(0.4% + 1dgt)
Page 5 of 12
DM-311
Test DC
Voltage
190mV
1.9V
19V
190V
1000V
99-08-26