Diagnostic Record; Flowcell Problems; Gas Problems - GE DigitalFlow GS868 Service Manual

Panametrics steam ultrasonic mass flowmeter
Hide thumbs Also See for DigitalFlow GS868:
Table of Contents

Advertisement

Chapter 3. Diagnostics

3.3 Diagnostic Record

Upon leaving the Diagnostic Menu via the
continue to show the last diagnostic parameter that was selected. To return to normal measurement mode, select a
channel to display (for a 2-Channel meter only) and then select the desired display parameter. See Chapter 3,
Operation, of the Startup Guide for detailed instructions.
The values for the diagnostic parameters immediately after initial installation of the meter and verification of proper
operation should be entered in Table 7 in Appendix A, Service Record. These values can then be compared to future
values to help diagnose any future malfunction of the system.

3.4 Flowcell Problems

If preliminary troubleshooting with the Error Code Messages and/or the Diagnostic Menu indicates a possible flowcell
problem, proceed with this section. Flowcell problems fall into two categories:

gas problems

pipe problems.
Read the following sections carefully to determine if the problem is indeed related to the flowcell. If the instructions in
this section fail to resolve the problem, contact the factory for assistance.
3.4.1 Gas Problems
Most gas-related problems result from a failure to observe the flowmeter system installation instructions, as described
in the Startup Guide. Refer to Chapter 1, Installation, of the Startup Guide to correct any installation problems.
If the physical installation of the system meets the recommended specifications, it is possible that the gas itself may be
preventing accurate flow rate measurements. The gas being measured must meet the following requirements:
1. The gas must be homogeneous, single-phase and relatively clean. Although a low level of entrained particles may
have little effect on the operation of the Model GS868, excessive amounts of solid or liquid particles will absorb or
disperse the ultrasound signals. This interference with the ultrasound transmissions through the gas will cause
inaccurate flow rate measurements. In addition, temperature gradients in the gas flow may result in erratic or
inaccurate flow rate readings.
2. The gas must not excessively attenuate ultrasound signals.
Some gases (i.e., high-purity carbon dioxide, hydrogen, nitrogen, etc.) readily absorb ultrasound energy. In such a
E1
case, an
error code message will appear on the display screen to indicate that the ultrasonic signal strength is
insufficient for reliable measurements.
3. The gas sound speed must not vary excessively.
The Model GS868 will tolerate relatively large changes in the gas sound speed, as may be caused by variations in
gas composition and/or temperature. However, such changes must occur slowly. Rapid fluctuations in the gas
sound speed, to a value that is considerably different from that programmed into the Model GS868, will result in
erratic or inaccurate flow rate readings. Refer to Chapter 2, Initial Setup, of the Startup Guide and make sure that
the appropriate sound speed is programmed into the meter.
28
[EXIT]
EXIT
key or the
option on the option bar, the display screen will
DigitalFlow™ GS868 Service Manual