Introduction Safety notes Description SITRANS F Installing/Mounting Electromagnetic flowmeters SITRANS FM MAG 5000/6000 IP67 Connecting Commissioning Operating Instructions Operating Service and maintenance Diagnostics and Troubleshooting Technical data Spare parts/Accessories Menu diagrams Factory settings Product documentation and support 7ME6910 (SITRANS MAG 5000)
Page 2
Note the following: WARNING Siemens products may only be used for the applications described in the catalog and in the relevant technical documentation. If products and components from other manufacturers are used, these must be recommended or approved by Siemens. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems.
Edition Remarks FW version 12/2019 4.09 ● BBL42 as default unit ● Improved operation without SENSORPROM ● Responsibility transfer to Siemens AG 12/2013 4.07 ● Customer defined unit ● Velocity value with unit ● Operational without SENSORPROM ● Signal suitability 01/2012 4.04...
Siemens’ products and solutions undergo continuous development to make them more secure. Siemens strongly recommends that product updates are applied as soon as they are available and that the latest product versions are used. Use of product versions that are no longer...
The content reflects the technical status at the time of publishing. Siemens reserves the right to make technical changes in the course of further development.
Page 10
Introduction 1.7 Notes on warranty SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
50% relative humidity from 40°C (104°F) ● Main supply voltage fluctuations up to ±10% of the nominal voltage (see Technical specifications (Page 67)) ● Overvoltage category II ● Pollution degree 2 SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
CSA Class I, Division 2, Groups A, B, C and D. Code T5 for an ambient temperature of +60 °C. FM Class I, Division 2, Groups A, B, C and D and Class I, Zone 2, Group IIC indoor/outdoor Type IP67 hazardous (classified) locations. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 13
● EN60079-31 is considered for installation in areas with combustible dust. ● When protective earth (PE) is connected, no potential difference between the protective earth (PE) and the potential equalization (PA) can exist, even during a fault condition. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 14
Safety notes 2.3 Installation in hazardous location SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
The main applications can be found in the following sectors: ● Water and waste water ● Chemical and pharmaceutical industries ● Food & beverage industry ● Mining and cements industries ● Pulp and paper industry SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
MAG 5000/MAG 6000 versions The transmitters are designed in various versions and offer high performance and easy installation, commissioning and maintenance. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
1. Check for mechanical damage due to possible improper handling during shipment. All claims for damage are to be made promptly to the shipper. 2. Make sure the scope of delivery, and the information on the type plate corresponds to the ordering information. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 18
Order number ④ Serial number ⑤ Power supply ⑥ Degree of protection ⑦ Ambient temperature ⑧ Conformity with country-specific directives ⑨ Place of manufacture Figure 3-1 MAG 6000 transmitter nameplate example SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 19
⑦ Software version and hardware version ⑧ Ambient temperature ⑨ Uncertainty ⑩ Certification number ⑪ Conformity with country-specific directives ⑫ Place of manufacture Figure 3-2 MAG 6000 CT system nameplate example SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 20
MAG 5000/6000 CT system sensor nameplate example Note The matched paired transmitter and sensor shall be mounted together At installation, please check that the system nameplates of transmitter and sensor have the same system serial number. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
No further assembling is necessary. The transmitter can be installed either compact on the sensor or remote. Figure 4-1 Compact installation Figure 4-2 Remote installation SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Reading and operating the flowmeter is possible under almost any installation conditions because the display can be oriented in relation to the sensor. To ensure optimum flow measurement, attention should be paid to the following: Vibrations Figure 4-3 Avoid strong vibrations SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 23
Figure 4-4 Medium and ambient temperatures Remote installation Cable length and type (as described in Cable requirements (Page 76)) must be used. For installation conditions for sensors, see relevant sensor Operating Instructions. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
System will not register flow if black plugs are not connected to connection board. 7. Fit supply and output cables through cable glands and connect to connection plate as shown in Electrical connection (Page 38). SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 25
11.Transmitter is ready to be powered up. NOTICE Exposing transmitter to direct sunlight may increase operating temperature above its specified limit, and decrease display visibility. A sunshield is available as accessory. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
(terminals 82, 0 and 83). 3. Remove SENSORPROM (to be mounted in transmitter terminal box). Ensure that the serial no. on the SENSORPROM label is identical to the sensor serial no. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 27
③ ④ 4. Connect electrode cable and coil cable in the corresponding terminals on the terminal blocks. 5. Remount terminal box lid. Note Tightening torque Tighten the bolts with 0.5 Nm. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 28
2. Ensure that correct SENSORPROM memory unit is mounted in wall/pipe mounting unit. ® 3. Fit M20 or ½" NPT cable glands for cables from bottom or sides of terminal box. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 29
8. Tighten cable glands to obtain optimum sealing. CAUTION Coil cable shield When remote mounted, power supply PE wire must be connected to PE terminal ( Coil cable shield must be connected to SHIELD terminal. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Setting of primary operating parameters is blocked during normal operation. When key is mounted, access to all menu items is gained. When key is removed, primary settings are blocked in accordance with requirements in authorisation. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Sealing of MAG 5000/6000 CT 4.5.2.1 Verification sealing Note The verification sealing is done at factory after initial verification. Under transmitter display frame On bottom of transmitter On side of transmitter SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
● DN 350 to 1200 (14" to 48") only in horizontal installation ● Compact or remote with max. 500 m (1640 ft.) cable ● Power supply 115 to 230 V AC and 12 to 24 V DC SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
● Compact or remote with max. 500 m (1640 ft.) cable Other restrictions may apply (see certificate) Turning transmitter/keypad Note Not allowed for MAG 5000/6000 CT Altering the orientation of the transmitter or keypad is prohibited for non-CT versions. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 34
Terminal box can be rotated ±90° in order to optimize viewing angle of transmitter display/ keypad 1. Unscrew the four screws in bottom of terminal box. 2. Turn terminal box to required position. 3. Retighten screws firmly. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 35
Installing/Mounting 4.6 Turning transmitter/keypad Keypad 1. Remove outer frame using a screwdriver. 2. Loosen the four screws retaining keypad. 3. Withdraw keypad and turn it to required orientation. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 36
Installing/Mounting 4.6 Turning transmitter/keypad 4. Tighten the four screws until a mechanical stop is felt in order to obtain IP67 enclosure. 5. Snaplock outer frame onto keypad (click). SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
A switch or circuit breaker (max. 15 A) must be installed in close proximity to the equipment and within easy reach of the operator. It must be marked as the disconnecting device for the equipment. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Wiring diagram Note Terminals 81 and 84 are only to be connected if special electrode cable with double screening is used, e.g. when empty pipe function or long cables are used. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 39
Fasten cables/wires in a way that they cannot get into contact with each other, not even in case of an error. Keep wire ends as short as possible. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
When the add-on module has been installed, the electrical connections are available on terminal rows 91-97. For more information Refer to the relevant BUS communication Quick Start or Operating Instructions available on an included DVD or on the internet, at: www.siemens.com/flowdocumentation (www.siemens.com/flowdocumentation). SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
All changed data will be stored in SENSORPROM memory unit. ® 4. Remove standard transmitter and remount Blind transmitter. 5. Fasten screws holding transmitter. New settings stored in SENSORPROM memory unit will be uploaded in blind transmitter. ® SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Operator menu Output Operator-active Product identity External input Operator-inactive Lock field symbols Ready for change Access to submenu Value locked RESET MODE: Zero setting of totalizers and initializa‐ tion of setting SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Access to a submenu in the setup menu is gained by pressing the lock key . Press the top up key briefly to return to the previous menu. Press longer (2 sec.) to exit the setup menu and return to operator menu No. 1. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Setting of measuring range, analog outputs and frequency output. Also individ‐ ual dimension-dependent setting of value, decimal point, unit and time. This menu is only visible if chosen as external digital input. Totalizer Setting of unit and decimal point. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 45
6. Press lock/unlock key to unlock settings. 7. Use select key and change key to change settings. 8. Press lock/unlock key to confirm new settings. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 46
10.Press top up key two times to exit setup mode. − − − − − − − Decimal point can be positioned and units set individually for flow rate in totalizer 1 and totalizer SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Note If "Text" is selected in line 2, this line functions as a heading for the value shown in line 3. Otherwise it shows the actual value of the reading selected. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
4. Press lock/unlock key to unlock language. 5. Use change key to select desired language. 6. Press lock/unlock key to confirm new language. 7. Press top up key two times to exit setup mode. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 49
Commissioning 6.7 Changing language SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 50
Commissioning 6.7 Changing language SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
High 22.6 mA For setting of error level, see Digital output / Relay output - Error level (Page 84). If current output is not used, it must be set to "Off". SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 52
Not available in MAG 5000, MAG 5000 CT and MAG 6000 CT. Note Cleaning If a cleaning unit is installed together with transmitter, relay output must always be used to operate this unit. It cannot be used for other purposes. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
If a SENSORPROM is not installed, check the sensor characteristics to see if they match the product label and the previous customer settings. Reset mode The reset mode is used for resetting/presetting totalizers or for restoring the factory settings. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 54
7. Use change key to select "manual". 8. Press forward key and then select key and change key to key in offset value. 9. Press lock/unlock key to start adjustment. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
● When choosing hot water, changing the output settings is not allowed and the output setting menus are not shown in display. ● When choosing cold water or other liquids, all output settings can be changed. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
When this is done, the decrease in measuring accuracy can be kept below 1% o.r. A too high frequency for the sensor used will cause a coil current alarm indication. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
● Seal integrity of the process connections, cable entries, and cover screws ● Reliability of power supply, lightning protection, and grounds NOTICE Repair and service must be carried out by Siemens authorized personnel only. Note Siemens defines flow sensors as non-repairable products.
Further information about our technical support is available on the Internet at Technical support (http://support.automation.siemens.com/WW/view/en/16604318) Service & Support on the Internet In addition to our documentation, we offer a comprehensive knowledge base online on the Internet at: Service and support (http://www.siemens.com/automation/service&support) SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
● Information about field service, repairs, spare parts and lots more under Services. Additional Support If you have additional questions about the device, please contact your local Siemens representative and offices at: Local contact person (http://www.automation.siemens.com/partner) Return procedure Enclose the bill of lading, return document and decontamination certificate in a clear plastic pouch and attach it firmly to the outside of the packaging.
Service and maintenance 8.4 Recalibration Recalibration Siemens AG offers to recalibrate the sensor. Please use the recalibration MLFB 9LA110-8Qxxx-xxxx, where xxx-xxxx indicate customer-specific configuration. Note For recalibration the SENSORPROM memory unit must always be returned with the sensor. ® SITRANS FM MAG 5000/6000 IP67...
The acceptance level for "error pending" can be individually configured to a particular application. The acceptance level is set in the basic settings menu (Page 44). SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 62
An output can also be selected to activate on a single error number. The alarm field, error output and error pending always operate together. Operator menu Error pending and status log are as default enabled ( ) in the operator menu. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Switch off and on A parameter is out of range. The parameter could not Active Active be replaced by its default value. The error will disap‐ pear at the next power-on SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 64
Empty display Minimum 1. No power supply Power supply Check MAG 5000/6000/6000 I for ben‐ ded pins on the connector 2. MAG 5000/6000/6000 I de‐ Replace MAG 5000/6000/6000 I fective SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 65
Pulse overflow Volume/pulse too small Change volume/pulse Pulse width too large Change pulse width Measuring approx. Missing one electrode connec‐ Check cables tion Loss of totalizer data Initializing error Reset totalizer manually SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 66
Reset totalizer or increase totalizer unit Empty pipe error mes‐ Empty pipe error Switch off MAG 5000/6000/6000I, wait sage when Empty 5 sec. and switch on again pipe set to off SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
± 0.1% of actual flow for V ≥ 0.5 m/s (1.5 ft/s) and con‐ ductivity ≥ 10 μS/cm 10.3 Output characteristics Current output Output characteristics Bidirectional mode Unidirectional mode 0 to 20 mA 4 to 20 mA Frequency output Bidirectional mode Unidirectional mode SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 73
Active Error relay output Bidirectional mode Unidirectional mode No error Error Limit switch (can be used as direction switch) 1 set point 2 set points Passive Digital output Passive Digital output SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 74
If 2 set points must activate 2 separate outputs, a single set point has to be selected individually for digital as well as relay outputs. Batch on digital output Unidirectional mode (forward flow only) SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
1.5 mm 0.25 mm Screen Double Color code Brown, blue, black Brown, blue, black Outside color Grey Grey Ext. diameter 7.8 mm 8.1 mm Conductor Flexible CU Flexible CU Isolation material SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
For Ex zone 1 installations only certified cable glands with protection type "e" can be used for the power supply and the coil cable. The cable glands must be approved for the actual temperature and cable dimension. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Ensure that your ordering data is not outdated. The latest ordering data is always available on the Internet: SIOS catalog (https://support.industry.siemens.com/cs/products? dtp=Catalog&mfn=ps&pnid=17318&lc=en-US) Spare parts Description Connection plate SENSORPROM memory unit ® Display unit Communication modules for MAG 6000 SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Spare parts/Accessories A.3 Sun shield Sun shield Description Sun shield SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Menu diagrams Transmitter menu overview The menu diagrams shown on the following pages apply to MAG 5000/6000 as well as MAG 6000 I. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Menu diagrams B.2 Basic settings Basic settings Note Locked or hidden menus Depending on your configuration some menus might be locked or hidden. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 81
Menu diagrams B.2 Basic settings − − − − SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 82
Menu diagrams B.2 Basic settings − − − When batch is selected on digital output or relay, Totalizer 2 is not shown because it is controlled by the batch function. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Do not combine differentiation and low alarm level together with passive output. The output will try to pull down the level to 1.3 mA at fatal errors which is not possible for passive output. Digital output - pulse SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Menu diagrams B.7 Digital output / Relay output - Error number Digital output - frequency Digital output / Relay output - Error level Digital output / Relay output - Error number SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
If cleaning unit is installed, relay outputs must always be used to operate cleaning. Relay outputs cannot be used for other purposes B.10 Digital output / Relay output - Batch Note This menu does not exist on MAG 5000/6000 CT SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 86
If digital output is set up for pulse, frequency or batch, then the relay output will be turned off if it has been set up for batch. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
(The display is locked during this time). The time cycle can be set at 1 to 240 hours. If the cycle is set at for example three hours, the transmitter will be active every three hours. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
In this case the power must be switched off/on to suppress the P40 error on the output. The frequency can be set to 44 Hz in the MAG 6000 SV transmitter only. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
9 is the best signal measured. If digital output is set to pulse (standard). If digital output is set to frequency. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
'Text' means that the text for the chosen measured value is shown. For example, if textis chosen in line 2 and flow velocity is chosen in line 3, the text "Flow velocity " is shown in line2 and the measured flow velocity is shown in line 3. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Menu diagrams B.19 Change password B.18 Add-on communication module Example: HART Note Burst mode is not available with HART communication B.19 Change password SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
1.5% 0 to 9.9% Empty pipe On, Off Velocity unit m, mm, cm, dm, ft , in per s, min, h, d, ? (customer unit) Error level Warning Fatal, permanent, warning SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 96
Correction factor 0.85 to 2.00 Sensor charac‐ teristics teristics (Page 88) Language Language English English, German, French, Danish, Swedish, Fin‐ Changing lan‐ nish, Spanish, Russian, Italian, Portuguese, Pol‐ guage (Page 48) SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
40 (1½) 4.979526 199.181 4.979526 199.181 US GPM 50 (2) 6.932434 277.2973 7.78051 311.2203 US GPM 65 (2½) 11.00577 440.2305 13.14907 525.9624 US GPM 80 (3) 17.62753 705.1008 19.91811 796.7241 US GPM SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
Page 99
US GPM 3000 (120) 308201 28009.84 1120393 US GPM * Factory setting sets Qmax to a metric unit (see previous table). The values here are converted to rounded off US GPM. SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
33628.27 m 1500 (60) 1.472622 l 38603.89 m 1600 (66) 1.675517 l 43922.64 m 1800 (72) 2.120576 l 55589.6 m 2000 (78) 2.617994 l 68629.13 m 2200 (90) 3.167773 l 83041.25 m SITRANS FM MAG 5000/6000 IP67 Operating Instructions, 12/2019, A5E02338368-AD...
– "Download as html5, only PC": Open or save the manual in the HTML5 view on your PC You can also find manuals with the Mobile app at Industry Online Support (https:// support.industry.siemens.com/cs/ww/en/sc/2067). Download the app to your mobile device and scan the device QR code. Product documentation by serial number Using the PIA Life Cycle Portal, you can access the serial number-specific product information including technical specifications, spare parts, calibration data, or factory certificates.
In addition to our technical support, Siemens offers comprehensive online services at Service & Support (http://www.siemens.com/automation/service&support). Contact If you have further questions about the device, contact your local Siemens representative at Personal Contact (http://www.automation.siemens.com/partner). To find the contact for your product, go to "all products and branches" and select "Products &...