Siemens SIPROCESS GA700 Quick Start Manual page 124

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Technical specifications
11.4 Technical specifications of analyzer modules
Table 11- 14 OXYMAT 7: Gas inlet conditions
Gas inlet conditions
Sample gas pressure
Reference gas pressure
Pressure loss between sample gas
inlet and sample gas outlet
Sample gas flow
Sample gas temperature and humidity
OXYMAT 7 standard version
OXYMAT 7 high-temperature
Table 11- 15 OXYMAT 7: Temperature of sample chamber and gas path
Temperature of sample
chamber and gas path
Gas path
Sample chamber
124
Standard devices with hoses
Tolerance within the valid pressure
range
Standard devices with hoses with
external reference gas pump
Tolerance within the valid pressure
range
Standard devices with pipes
Tolerance within the valid pressure
range
High-pressure connection
Low-pressure version with external
reference gas pump
Without vibration compensation
With vibration compensation
500 to 1 500 hPa (absolute)
± 150 hPa
700 to 1 200 hPa (absolute)
± 150 hPa
500 to 3 000 hPa (absolute); short-term < 5000 hPa (abs.)
± 150 hPa
2 000 hPa above sample gas pressure
(within permitted reference gas pressure range
2500 to 5000 hPa, abs.)
100 hPa above sample gas pressure
150 hPa above sample gas pressure
< 100 hPa at 1 l/min
18 to 60 l/h (0.3 to 1 l/min)
0 to 60 °C with relative humidity < 90%, avoid condensation
in the gas path (moisture content results in measurement
errors)
0 ... 130 °C with relative humidity < 90%, avoid condensa-
tion in the gas path (moisture content results in measure-
ment errors)
OXYMAT 7
5 K above ambient temperature
72 °C
Compact Operating Instructions, 05/2018, A5E31805656-06
OXYMAT 7 high-temperature
72 ... 130 °C, adjustable
72 ... 130 °C, adjustable
Quick Start