Command: CN - Get number of calibration
Format:
Description:
Example:
Query:
Command:
Format:
Description:
Query:
Command:
Format:
Description:
Example:
Query: There is no associated query command.
points required
CN?
Returns the minimum and maximum
calibration points required to carry out the
specified calibration procedure.
IU = 0
set-up input as mbar
PP = 123 PIN entered put instrument
into calibration mode
CT = 1
perform two point calibration
CN?
Reply: CN=1,2 requires between 1 and 2
calibration points
This is a query only command.
CA - Calibration Accept
CA
Accepts the calibrated data. If the correct
number of calibration points have been
entered with the CP command then the
calibration coefficients are calculated and
stored in the instrument's nonvolatile
memory. The instrument returns to
measurement mode.
There is no associated query command.
CX - Calibration abort
CX
Aborts the calibration procedure.
IU = 0
set-up input as mbar
PP= 123
PIN entered put instrument
into calibration mode
CT = 1
perform two point calibration
CP = 200
applied pressure 200 mbar
CX
abort calibration