Chapter 3. Inserting T9 Transducers into the Pipe
3.2
Inserting Transducers with the Simple Holder (cont.)
6. Use the equation below to calculate the path length (P) based on the pipe inside diameter and transducer face angle:
ID
where:
= pipe inside diameter and
A
= nozzle angle
7. Record the axial dimension (L). The L dimension is equal to the pipe inside diameter.
8. Refer to your flowmeter Startup Guide or User's Manual to make transducer electrical connections.
You have completed transducer installation.
18
P = ID/cos (A)
P
L
Model T9 Ultrasonic Flow Transducer Installation Guide