Download Print this page

Maintenance - Sealey SCMA1 Manual

Carbon monoxide alarm

Advertisement

7.11.
END OF ALARM LIFE WARNING
7.11.1.
When the alarm is at the end of its life, it gives three "beep" audible alarms every 60 seconds together with a LED visual alarm, see
fig.8.
7.12.
CALIBRATION
Calibration can only be carried out by a professional organisation, please DO NOT calibrate it yourself.
7.12.1.
7.12.2.
Calibration is carried out on a gas concentration of 400 PPM, this will not affect the unit's accuracy, every unit has been calibrated
before delivery.
7.12.3.
Press the TEST button down until all 3 LEDs are on, then release the TEST button, the green and yellow LEDs will turn off, only the
red LED will stay on. It signifies that the alarm is now in Calibration Mode.
7.12.4.
The unit will now give 5 minutes in which it can be calibrated.
7.12.5.
If the unit cannot detect any change of CO concentration within 5 minutes it will automatically exit Calibration Mode.
7.12.6.
After calibrating on CO gas, the 3 LEDs will flash 3 times at the same time to indicate the Calibration was successful.
NOTE:
1. The unit has a minimum CO concentration for calibrating (250 PPM minimum). If the calibration is carried out on a concentration
lower than 250 PPM, the unit will not change its previous calibration values.
2. When the CO concentration on the LCD is bigger than 30 PPM, the unit will stop entering into Calibration Mode to prevent
mis-operation.
3. If the user accidentally enters into Calibration Mode by chance, just press the TEST button to exit the Calibration Mode.
7.13.
WHEN THE CO ALARM RAISES AN ALARM
7.13.1.
Immediately move outdoors, open all doors/windows to ventilate the area and allow the carbon monoxide to disperse.
7.13.2.
Where possible to, turn off any fuel burning appliance.
7.13.3.
Evacuate the building. DO NOT re-enter the premises or move away from the open door/window until emergency services have
arrived, the premises have been aired out and your alarm has returned to its normal condition.
7.13.4.
Seek medical help immediately for anyone suffering from symptoms of headache, drowsiness, nausea etc.
DO NOT use the appliances again until they have been checked by a qualified technician, the fault located and repaired.
8

8. MAINTENANCE

8.1.
Replace the entire unit every 5 years.
8.1.
Your detector will alert you to potentially hazardous CO concentrations in your home when maintained properly.
8.2.
To maintain your alarm in proper working order it is recommended that you:
• Test the alarm by pressing the TEST button at least once a month.
• Clean your CO alarm regularly to prevent dust build up. This can be done with a vacuum cleaner with the brush attachment once a
month. Clean gently around the grilles and sides.
• Never use cleaning solutions on the alarm, simply wipe with a damp cloth.
DO NOT paint the alarm.
8
9. GAS SELECTIVITY
9.1.
This alarm possess conformity to selectivity which international standards require and the table below shows the selectivity to
gases defined in UL2034 and EN50291.
TEST GAS
CO
Hydrogen
Methane
Iso-Butane
Carbon dioxide
Carbon di-sulfate
Hydrogen sulfide
Nitrogen Oxide
Nitrogen Dioxide
Ammonia
Ethyl acetate
Di-chloromethane
Heptane
Toluene
IPA
Ethanol
Hexa-methyl di-siloxan
Exposure time *:30 minutes **:40minutes ***:2hours
© Jack Sealey Limited
RELATIVE SENSITIVITY (CO IS 100)
100
40
0
0
0
0
0
0
Less than 5
0***
0***
0***
0***
0***
0***
Less than 2*
0**
Original Language Version
SCMA1 Issue 2 23/06/2020

Advertisement

loading

This manual is also suitable for:

5051747923232