Operation - GE FineLINE 35 oligo Operating Instructions Manual

Table of Contents

Advertisement

4 Operation

FineLINE 35 oligo column Operating Instructions 28-9649-57 AA
WARNING
Hazardous substances. When using hazardous chemical and
biological agents, take all suitable protective measures, such as
wearing protective glasses and gloves resistant to the substances used.
Follow local and/or national regulations for safe operation and
maintenance of the system.
WARNING
The design pressure must not be exceeded.
See Section 7.1 Specifications‚ on page 35 for pressure limit.
WARNING
Emergency Stop. When the column is fed by an external pump or
system, the Emergency Stop for that system shall be used in case of an
emergency. Refer to the Operating Instructions for the used pump or
system regarding Emergency Stop before using the column.
CAUTION
Incorrectly fitted tubing may loosen, causing a jet of liquid to spray out.
This is especially dangerous if hazardous chemicals are in use. Connect
the tubing by first inserting the tubing fully, then tightening the
connector fingertight. PEEK tubing should be tightened a further 1/4
turn using the supplied spanner or wrench. Do not tighten Teflon tubing
further as this will damage the end of the tubing.
NOTICE
No other solvents than those specified in the Operating Instructions
should be used as they could possible damage the system.
NOTICE
The column material is not compatible with tetrahydrofurane, THF.
Operation 4
19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents