4 Installation
4.4 ÄKTAprocess Setup

4.4.5 Connect sample pump

4.4.5
Connect sample pump
Overview
If ÄKTAprocess is delivered with the sample pump option, connect the sample pump
with the enclosed tubing using TC-connectors. Refer to the Section 3.5 Flowchart, on
page 72 for sample pump connection location.
The sample should be fed via the inlet air sensor.
Precaution
Preparations
Follow the instruction below to prepare for connecting the sample pump.
Step
1
2
3
100
WARNING
Read and understand all precautions in Sample pump, on page 24
before attempting any actions described in this section.
Action
Position the sample pump trolley beside the system.
Note:
The best position is adjacent to the right-hand corner of the system when
facing the operator console. Make sure that the sample pump trolley is posi-
tioned so that it does not block the column valves and the emergency stop
switch.
Log out of UNICORN, turn off the computer and shut down the system. See
Procedures after usage.
Note:
The system must be re-started after the sample pump has been connected
in order for the monitoring of the sample pump pressure sensor to be activat-
ed.
Disconnect the compressed air supply from the system.
ÄKTAprocess Operating Instructions 29-0252-49 AA