Point Of Use (Pre-Cleaning); Containment And Transportation; Manual Cleaning And Disinfection - Hitachi EUP-L53 Instruction Manual

Linear array probes
Table of Contents

Advertisement

4.1

Point of use (Pre-cleaning)

Pre-cleaning should be done immediately after each use.
The procedure is as follows:
1) Remove the probe cover and any accessories from the probe.
2) Clean the probe of all patient's blood or fluid with running tap
water until the surface of the probe looks visually clean.
3) Wipe the whole surface of the probe with gauze pad and remove
superficial visible impurities.
4.2

Containment and transportation

Putting the contaminated equipment into exclusive
shock and damage proof container for transportation
is recommended. It is recommended that instruments are reprocessed as
soon as possible and not later than 4 hours after usage.
4.3

Manual Cleaning and disinfection

Prepare following items before manual cleaning
and disinfection:
a) Detergent: Cidezyme® (Johnson & Johnson,
#2258) or another cleaning agent with
approved material compatibility for this
medical device.
b) Disinfectant:
Johnson, # 20391) or another disinfectant
with approved material compatibility for
this medical device.
c) Two tanks, one for cleaning and one for
disinfection - optional:
1 additional tank for rinsing with deionized/tap water (sufficient
size for immersion of the submergible part of the probe at full
length)
d) Soft, fluff free cloth or single use towel
e) Personal protective equipment (gloves, water repellent protective
skirt, face protection mask or protective glasses, see also
instructions of the manufacturer for the detergent and the
disinfectant)
Cidex®
OPA
(Johnson
-13-
Point of use
(Pre-cleaning)
Containment and
transportation
Manual cleaning and
disinfection
Manual Cleaning
Rinsing after manual
cleaning
&
Manual Disinfection
Rinsing after manual
disinfection
Q1E-EP0348