Explosion protection in accordance with FM (USA)
Special conditions for use
For type of protection "db": When you use the device as partition wall device in zones that require EPL Ga, isolate the non-
intrinsically safe circuit safely from ground, e.g.: through a SELV circuit.
For type of protection "ec": Isolate the non-intrinsically safe circuit safely from ground, e.g.: through a SELV circuit that
ensures transient protection up to 140% of the nominal voltage.
Mount the device so that the enclosure is not exposed to any intense electrostatic charging processes.
To prevent electrostatic charging in hazardous areas, clean the outside of the enclosure using a cloth moistened with water.
Explosion protection in accordance with cCSAus (Canada/USA)
CSA18CA70163103X
Marking (XP/DIP) or (IS); (NI)
Further certificates for explosion protection
Explosion protection in accordance with NEPSI (China)
GYJ19.1058X
Explosion protection in accordance with INMETRO (Bra‐
sil)
BRA-18-GE-0035X
Explosion protection in accordance with EAC (Russia)
TC RU C-DE.AA87.B.01202
Explosion protection in accordance with IECEx
IECEx BVS 18.0038X
SITRANS P320/P420
Compact Operating Instructions, 01/2021, A5E38874562-AC
The permissible operating values can be found in the certificate.
According to CSA
Ex ia/db IIC T4/T6 Ga/Gb
Standards
Ex ia IIC T4/T6 Ga
Ex ia IIIC T120°C Da
Ex tb IIIC T120°C Db
Ex ec IIC T4/T6 Gc
Class I, DIV 1, Gr. A-D (Explosion Proof)
Intrinsically Safe (IS) for Class I, DIV 1, Gr. A-D
(Intrinsically Safe (IS))
Class II, DIV 2, Gr. E-G; Class III
Class I, DIV 2, Gr. A-D (NI)
According to US
Class I, Zone 0/1, AEx ia/db IIC T4/T6 Ga/Gb
Standards
Class I, Zone 0, AEx ia IIC T4/T6 Ga
Zone 20, AEx ia IIIC T120°C Da
Zone 21, A/Ex tb IIIC T120°C Db
Class I, Zone 2, AEx ec IIC T4/T6 Gc
Class I, DIV 1 Gr. A-D (Explosion proof (XP))
Class I, DIV 1 Gr. A-D (Intrinsically Safe (IS))
Class II, DIV 2, Gr. E-G; Class III
Class I, DIV 2, Gr. A-D (NI)
The permissible operating values and markings can be found in the