Combination operation
3.3 Gas connections
Retrofitting example "Suction operation, serial" operating mode for OXYMAT 7/ULTRAMAT 7
When you are using the ULTRAMAT 7 and the OXYMAT 7 analyzer module in serial suction
operation, you must convert the sample gas restrictor of the OXYMAT 7 analyzer module:
Sample gas inlet → Sample gas outlet.
MGE, VGE
MGA/VGA
Figure 3-7
32
Sample gas inlet (MGE), Reference gas inlet (VGE)
Sample gas outlet (MGA)/reference gas outlet
Remove sample gas restrictor from the sample gas inlet of the additional OXYMAT 7
analyzer module
Install the clamping screw into the sample gas inlet of the additional OXYMAT 7 analyzer
module
Install the sample gas restrictor into the sample gas outlet of the additional OXYMAT 7
analyzer module
Install additional OXYMAT 7 analyzer module
Target configuration: ULTRAMAT 7 and OXYMAT 7 in serial suction operation
Retrofitting of the sample gas restrictors: Example "Pressure operation, serial"
Compact Operating Instructions, 05/2015, A5E31805656-03