Option Strings Check - GE LOGIQ e R4 Series Basic Service Manual

Table of Contents

Advertisement

GE
Direction 5370626-100, Revision 18
8-3-4

Option Strings Check

NOTE:
After the system software loading completion, please check the option strings to ensure
that the options are activated and working.
1.) Reboot the system.
2.) Select Utility-Admin-System Admin
3.) Ensure that all the installed Option keys are displayed and the status of Options are valid.
The status "valid" means the option keys are activated and working.
The status "disabled" means the option keys are not activated and not working. Check if the option
is installed and if serial number and option key are correct.
8-20
LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Figure 8-26 Option strings
Section 8-3 - Loading the System Software