Section 6
Technical data
36
Table 12:
Degree of protection of BIO-Tester Relion
Description
Closed lid
Open lid
Table 13:
Environmental conditions
Description
Application field
Operating temperature range
Transport and storage temperature range
Humidity range
Table 14:
Ratings of output connector terminals
Description
PSM1001/1002/
Rated voltage
1003 connector
Rated power
BIO1001/1003
Binary inputs
connector and
BIO1002/1004
connector
Binary
outputs
LHMI 48 VDC
Rated voltage
connector
Rated power
Table 15:
Main power switch and main's wall plug connector
Description
Rated frequency
Rated voltage
Rated power
Rated input current
Fuse type
®
REX640
Value
IP 67
IP 20
Value
Indoor use
+5...50°C (+41...122°F)
-25...+70°C (-13...+158°F)
Maximum relative humidity 80% for temperature
up to 31°C decreasing linearly to 50% relative
humidity at 40°C
Value
48 or 110 VDC
Max. 100 W
Rated
12...200 VDC
voltage
Rated power
Max. 15 W (200 VDC)
SW fuse
~65 mA
Rated
48 VDC
voltage
Rated power
Max. 100 W
48 VDC
Max. 100 W
Value
50/60 Hz
100...240 VAC
Max. 230 W
1 A
T2A/250 V
2NGA000105 B