8.2 System software - Installation/Upgrade procedure
Introduction
There are two possibilities to update the system software:
1.
via the
-
Software parts to be upgraded (e.g., Ultrasound Application Software, Service Software, EUM,
MS patches, etc.) and installation time depend on contents of the used System DVD.
-
If the currently installed software has to be upgraded to a newer version, a new software specific
"Permanent key" is required. Please contact your local distributor or GE service representative.
2.
by downloading and installing a
-
Software parts to be upgraded (e.g., Ultrasound Application Software, Service Software, EUM,
MS patches, etc.) and installation time depend on contents of the SW update package.
-
PRECONDITION: InSite connection has to be configured and checked out. Operation see:
Section 3.13.4 "InSite ExC Configuration" on page 3-77
Caution
Disconnecting ALL external USB devices (except DVD drive) is NECESSARY. Re-installation of any
previously attached printer has to be done after the upgrade procedure.
Installing the Bluetooth Printer and its connection set is NOT possible by the user.
Note
Manpower
One person ~ 1 hour (depends on contents of System DVD resp. SW update package, peripherals, etc.)
Tools
System DVD
Voluson E-Series Service Manual
KTD106657 Revision 2
FMI from DVD
button in the System Setup
Software Update Package
Replacement Procedures
Service page
via active InSite connection
.
8-3