Using Data Table 3 - GE MG-101 User Manual

Field moisture calibration system
Table of Contents

Advertisement

2.2.2 Using Data Table 3

For calibration points found in
1.
Fully open the vent valve by turning it counter-clockwise.
CAUTION!
Make the following adjustments slowly, being careful not to overshoot the target value. NEVER
ALLOW THE SYSTEM TO GO TO A VALUE WETTER THAN THE TARGET VALUE. If you overshoot a
value in the wetter direction, you will have to allow the system to equilibrate (dry down) again
before you can proceed.
2. In
, look below the body of the table for Notes 1 and 2. Adjust both the valves on the
Data Table 3
DILUTION
and the
DRY GAS SECOND DILUTION
IMPORTANT: Be sure to read the steel ball or glass ball (
3. Consult
Data Table 3
headed
WET GAS
. Adjust the
4. By referring to the desired dew/frost point temperature in
headed
WET GAS DILUTED
remain fully open) to obtain the value indicated in
After stabilizing, the system output will be at the selected dew/frost point temperature and it is ready for use. The
stabilization period depends on the dew/frost point temperature you are using. In general, the following times are
required:
For dew/frost point temperatures of -60°C (-76°F) and drier, allow 2 hours
For dew/frost point temperatures of -60°C (-76°F) to -40°C (-40°F), allow 1 hour
For dew/frost point temperatures of -40°C (-40°F) and wetter, allow 30 minutes
After you have completed the calibration or other calibrated moisture operation at this dew/frost point temperature,
either shut down the system (see "System Shutdown" on page 12), or repeat the steps in this section for the next
calibration point.
MG-101 User's Manual
, refer to Figure 6 on page 9 and proceed as follows:
Data Table 3
for the desired dew/frost point temperature and find the corresponding value in the column
WET GAS
valve to obtain the value indicated in the table.
. Adjust the vent valve (but not the valve on the
rotameters to the values listed.
or
), as indicated in the table.
S
G
, find the corresponding value in the column
Data Table 3
.
Data Table 3
Chapter 2. Setup and Operation
WET GAS DILUTED
rotameter, which must
DRY GAS FIRST
11