Performance Tests

SYNC Out and MARKER Out Amplitude Checks

4-40
These procedures check the amplitude of the SYNC Out and MARKER Out
signals.
Electrical
Auxiliary Outputs, Sync, Marker 1, Marker 2, Amplitude, on page 1-10.
Characteristic
Checked
Equipment
One 50 W coaxial cable
Required
One oscilloscope
Prerequisites
The AWG2021 meets the prerequisites listed on page 4-7.
1. Install test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2021 CH1 SYNC Out
connector through the coaxial cable to the CH1 vertical input connector
on the oscilloscope (see Figure 4–16).
AWG2021
AWG2021

Figure 4-16: Initial Test Hookup

b. Set oscilloscope controls:
Vertical
. . . . . . . . . . . . . . . . . . . . . .
CH1 Coupling
CH1 Scale
CH1 Input Impedance
Horizontal
Sweep
. . . . . . . . . . . . . . . . . . . .
Trigger
Source
. . . . . . . . . . . . . . . . . . .
Coupling
. . . . . . . . . . . . . . . . . .
Slope
. . . . . . . . . . . . . . . . . . . .
Level
. . . . . . . . . . . . . . . . . . . .
Mode
. . . . . . . . . . . . . . . . . . . .
Oscilloscope
TDS
540
CH1
. . . . . . . . . . . . .
DC
. . . . . . . . . . . . . . . .
200 mV/div.
50 W
. . . . . . .
50 ns/div.
CH1
DC
Positive
500 mV
Auto
AWG2021 Service Manual