Hitachi S21 Instruction Manual page 16

Table of Contents

Advertisement

see also instructions of the manufacturer for the detergent
and the disinfectant)
Manual Cleaning:
Prepare the detergent solution in a tank with cold water (please follow
the instructions of the detergent manufacturer regarding application,
dilution and contact time).
1) The temperature of the detergent solution should be between 15-30°C,
concentration is 1.6%. Please note the minimum contact time of the
detergent in the manufacturer's instruction. If a differing
detergent
is
compatibility for the medical device.
2) Immerse the immersible part of the probe without connector into the
diluted detergent solution (see Fig 3). Wipe the immersible part
of the probe under the surface of the detergent solution with a soft
cloth to remove all visible soil. Be sure that all grooves of the
probe are implemented during the cleaning process.
3) The immersible part of the probe should be left in the detergent
solution according to the specified contact time of the detergent
manufacturer.
4) Wipe the un-immersible parts of the probe with a soft cloth dipped
with the detergent solution.
5) Rinse the probe with running tap water for 1 minute. (alternatively:
immerse the immersible part of the probe in a tray filled with
deionized water/tap water (see Fig.3) for 5 min.)
6) Visually check the outer surface of the probe for cleanness. If
necessary, use magnifying glass for visually check. If there is still
soil visible, repeat all above steps.
used,
please
- 10 -
also
note
the
approved
material
Q1E-EP1377