Status/Error Reporting

Parameter/Operation Mnemonic
Status Byte
Status Byte Mask
Condition Byte
Self-test
Error Number
Last Error Number
Operation Complete
Identi er

Universal Commands

H-10 Remote Operation
Comment
Returns 3-digit integer (000-191)
Returns 3-digit integer (000-191)
Returns 2-digit integer (00-63)
Executes self-test, returns 0 for pass
and 1 for fail
Returns 3-digit integer representing error
code
Returns 3-digit integer for last active error
This is a destructive readout.
1 if no further commands to interpret and
execute in the input bu er
0 if further commands in the input bu er
Returns 56-character string identifying currently
installed rmware, manufacturer, model no. and
serial number.
j
Returns 26 character string identifying currently
installed rmware, manufacturerer, model no. and
serial number of channel A or B optical head.