Note
3-18
Test Descriptions
All manuals and user guides at all-guides.com
Purpose
Description
Test Settings
Settings for Amplitude Accy. INPUT 1 (10 dB AT and
Function
Band A
RF FREQ
9 kHz to 150 kHz 150 kHz to 30 MHz 30 MHz to 50 MHz
QPA lter
200 Hz
RES BW
3 kHz
RFPS (atten)
0 dB
SA (atten)
10 dB or 0 dB
Scale
Linear
0 dB AT)
Band B
9 kHz
100 kHz
0 dB
10 dB or 0 dB
Linear
HP 8572A EMI Receiver User's Guide
Band C
120 kHz
1 MHz
0 dB
10 dB or 0 dB
Linear