Final Steps - Siemens SIREMOBIL Iso-C Installation And Setting Instructions Manual

Hide thumbs Also See for SIREMOBIL Iso-C:

Advertisement

Installing a Video Printer
Switch the SIREMOBIL Iso-C off and on again to activate the new configuration.
For Memoskop C-E (without keyboard)
Switch on the system.
Connect the service PC to the service interface of the SIREMOBIL Iso-C.
Install and start-up the MEMOSKOP service program on the service PC.
Program "Hardcopy release" on SIREMOBL Iso-C systems with one monitor.
Program "Hardcopy release" on SIREMOBIL Iso-C systems with two monitors.
If there is an area dose product measurement device present (Diamentor), activate
"Dose on hardcopy output" as well.
Transfer the parameters to the MEMOSKOP.
Exit the MEMOSKOP service program.
Function test
Call up a fluoroscopic image on monitor A or monitor B.
Release an exposure at the operating panel of the basic system and at the monitor
trolley keyboard.


Final steps

Switch off the SIREMOBIL Iso-C.
Test the protective conductor using the protective conductor meter 44 15 899 RV 090.
Refer to ARTD - 002.731.17...
Changes as compared to previous versions
Paragraph on page 3 and 4 "Power cable" changed.
TD SD 24 / Arnold
Siemens AG
Medical Engineering
The video printer should print an image in both cases.