This document describes the replacement of the cable harness for MAMMOMAT 3000 (CE). Instructions are included concerning service work on MAMMOMAT 300 that differs signifi- cantly from work done on Mammomat 3000. Work performed on the Mammomat 300 may still deviate slightly from these instructions. Siemens AG Register 5 RXB7-230.091.05...
Remove the edge protection strips only when you have reat- tached the side covers or if a height adjustment is necessary. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 2 of 2 Siemens AG Rev.
X808, connected to control panel (10)inside the shaft cover. CAUTION Be careful not to damage connectors X807 and X808 when remov- ing the covers. Siemens AG Register 5 RXB7-230.091.05 Page 3 of 8 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
Fig. 4 Position of the protective conductors and the cable guide • Remove cable guide (1). • Remove earth screw (2) and ground cables (3). MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 4 of 8 Siemens AG Rev. 03 07.
Page 11
Pull out connectors X873 and X874, which are located behind the compression unit. • Loosen and remove screw 3, and remove the compression unit. Siemens AG Register 5 RXB7-230.091.05 Page 5 of 8 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
Remove the connection wires for the fan. • Remove the 4 screws (1/Fig. 6) and remove the fan unit (2/Fig. 6). Fig. 6 Removal of fan unit MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 6 of 8 Siemens AG Rev.
3 - 1 Remove the connectors at the MAMMOMAT 300/300 CE swivel X897 X855 X856 X860 X896 X870 X895 X894 4 emergency stop button cables (backside) RIGHT SIDE LEFT SIDE Fig. 1 • Remove all connectors in accordance with Fig. 1 and the list which follows (including the protective conductors).
• Eject wire #2 from connector X860. • Remove the four wires from the STOP switch. • Remove the wire with connector X893. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 2 of 16 Siemens AG Rev. 03 07. 98...
Page 17
5 protective conductors swivel arm frame 2 fan connection wires on the fan *1 MAMMOMAT 3000/ 3000 CE only Siemens AG Register 5 RXB7-230.091.05 Page 3 of 16 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
Lift up the X-ray tube and remove the high voltage cable (1) by unscrewing the sleeve nut (2). • Secure the X-ray tube again with the mounting screws. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 4 of 16 Siemens AG Rev.
Remove the component parts of the high voltage cable plug. • Cut the cable ties from the cable harnesses being removed. • Remove the cable from the swivel arm. MAMMOMAT 300/300 CE only - Removing the corrugated hose Fig. 5 R803 Fig. 6 •...
- The braided grounding wire should remain attached to the swivel arm. - Unfold the shielding and remove the semicircular cable clamp. • Pull the lead-through ring off the cable harness. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 6 of 16 Siemens AG Rev.
Remove only X893 (WING 1) Fig. 8 • Pull all cables out of the swivel arm as shown in Fig. 8. Siemens AG Register 5 RXB7-230.091.05 Page 7 of 16 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
Removing de cable harness All Mammomats: Removing the cable harness from the left side of the stand NOTICE for MAMMOMAT 300/ MAMMOMAT 300 CE: Some of the connectors below are omitted All connectors on D801, except flat cable to D802...
*1 omit for MAMMOMAT 3000 CE. MAMMOMAT 3000 CE only • Remove connector X802 in the base of the stand (transformer T802). Siemens AG Register 5 RXB7-230.091.05 Page 9 of 16 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
All Mammomats: Disconnecting the cable harness at the back wall of the stand NOTICE For MAMMOMAT 300/ MAMMOMAT 300 CE: Some of the connectors below are omitted. X887 Lift motor (Note the color before disconnecting X887) X888 and X826 Swivel motor...
- Remove the 3 screws show in (2/Fig. 11) and remove the swivel motor (3/Fig. 11). Fig. 11 Siemens AG Register 5 RXB7-230.091.05 Page 11 of 16 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
3 - 12 Removing de cable harness MAMMOMAT 300 CE and MAMMOMAT 3000 CE only Remove the screws from the connection device (1/Fig. 12) on the base of the stand and disconnect the following connectors: • X885 connection cable for the stereo evaluation unit.
Fig. 13 Fig. 14 • Loosen the hose clamp (1/Fig.13) for the corrugated hose on the lifting carriage.f for MAMMOMAT 300 CE and MAMMOMAT 3000 CE NOTICE The braided grounding wire remains attached to the lifting carriage. • Remove the angle bracket (1/Fig.14) for the corrugated hose.
X1 Serial interface. • X10, X11 Iontomat cable. • Protective conductor. Pull back all cables that lead from the generator to the base of the stand. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 14 of 16 Siemens AG Rev. 03 07.
Removing de cable harness 3 - 15 MAMMOMAT 300 / MAMMOMAT 3000 only Fig. 16 • Remove the cable bracket (1/Fig.16) from the base of the stand. All Mammomats • Withdraw the generator cable into the stand. • Remove the cable harness.
(for harness 62 40 175). • The conductors at X885 are connected as follows: white brown green Fig. 17 MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 16 of 16 Siemens AG Rev. 03 07. 98...
Route the cable (2/Fig. 2) in the lower part of the stand and secure it with cable ties. Fig. 2 Siemens AG Register 5 RXB7-230.091.05 Page 1 of 14 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
Page 32
• Route the branch with the connector for D806 on the lifting carriage as shown in 1/Fig. 3 and secure it with cable ties. Fig. 3 MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 2 of 14 Siemens AG Rev.
Connect X886 (inside of stand) X801 *1 not for MAMMOMAT 3000 CE Fig. 4 NOTICE Some connections may not be present on MAMMOMAT 300 / MAMMOMAT 300 CE. On MAMMOMAT 3000 CE: Plug X802 into the receptacle in the stand. Siemens AG Register 5 RXB7-230.091.05...
• Attach the cable bracket to the base of the stand (1/Fig. 6). Fig. 6 MAMMOMAT 300 CE / MAMMOMAT 3000CE only Connect the wires on the multi-purpose connection device in the stand: • Attach plug X885 using the screws and connect the wires to the stereo evaluation unit (if used).
Page 36
Installing the cable harness • Secure the angle bracket (2) to the corrugated hose with the screws. MAMMOMAT 300 CE / MAMMOMAT 3000 CE only Make sure that the shielding is folded over the lead-through ring NOTICE and that the semicircular cable clamp is in position. Also make sure that the braided grounding wire is between the hose clamp and the shielding.
Also make sure that the braided grounding wire is between the hose clamp and the shielding. • Install the stereo switch with the screws. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 8 of 14 Siemens AG Rev.
Divide the cable harness and insert it. Route the cable harness so that it is free of coils and kinks. • Secure the cable harness with cable ties. Siemens AG Register 5 RXB7-230.091.05 Page 9 of 14 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
Secure the cable with cable ties. • Reinsert wire 2 into connector X860. • Insert all cables as shown in the Fig. and connect as many plugs as possible at this point. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 10 of 14 Siemens AG Rev.
(light barriers for measuring the +/- 10° tilt of the tube.) • Reinstall the rubber bellows over WING 1 / WING 2 (refer to Pos. 7 in Fig. 2 - 8) with the screws. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 12 of 14 Siemens AG Rev.
Check the system functions according to the adjustment instructions (e.g. kV measurement, Iontomat test on both wings if fitted etc.). • Attach the generator cover. Siemens AG Register 5 RXB7-230.091.05 Page 13 of 14 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
Page 44
4 - 14 Installing the cable harness This page intentionally left blank. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 14 of 14 Siemens AG Rev. 03 07. 98 TD SD 24 Medical Engineering...
Page 45
Text changed 4-13 Text changed TD HSC 24 / Guggenmos TD SD 24 / Gross SMS Iselin / O’Donnell Siemens AG Register 5 RXB7-230.091.05 Page 1 of 2 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
Page 46
5 - 2 Changes to previous version This page intentionally left blank. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 2 of 2 Siemens AG Rev. 03 07. 98 TD SD 24 Medical Engineering...