MAMMOMAT 300 / 3000
Repair Instructions
Replacing the cable harness
Register 5
Print No.: RXB7-230.091.05.03.02
Replaces: RXB7-230.091.05.02.02
SP
© Siem ens A G 1 997
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model
_or_ design,_are_
reserved.
English
Doc. Gen. Date: 07. 98
65 03 093

Advertisement

Table of Contents
loading

Summary of Contents for Siemens MAMMOMAT 300

  • Page 1 MAMMOMAT 300 / 3000 Repair Instructions Replacing the cable harness © Siem ens A G 1 997 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights,...
  • Page 2 0 - 2 Revision Chapter Page Revision MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 2 of 4 Siemens AG Rev. 03 07. 98 TD SD 24 Medical Engineering...
  • Page 3: Table Of Contents

    Removing the cable harness at the generator ....3 - 14 MAMMOMAT 300 / MAMMOMAT 3000 only ..... 3 - 15 All Mammomats .
  • Page 4 MAMMOMAT 300 CE / MAMMOMAT 3000 CE only ....4 - 6 MAMMOMAT 300/ CE ....... . . 4 - 7 MAMMOMAT 3000/ CE.
  • Page 5: Systems Affected

    This document describes the replacement of the cable harness for MAMMOMAT 3000 (CE). Instructions are included concerning service work on MAMMOMAT 300 that differs signifi- cantly from work done on Mammomat 3000. Work performed on the Mammomat 300 may still deviate slightly from these instructions. Siemens AG Register 5 RXB7-230.091.05...
  • Page 6: Safety Information

    Remove the edge protection strips only when you have reat- tached the side covers or if a height adjustment is necessary. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 2 of 2 Siemens AG Rev.
  • Page 7: Removing The Stand Covers

    Remove screws (2) and inner plastic strip (3). • Remove screws (4) and front covers (5). • Remove screws (6) and rear side covers (7). Siemens AG Register 5 RXB7-230.091.05 Page 1 of 8 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
  • Page 8: Mammomat With Single Rear Cover

    • Remove screws (2) and inner plastic strip (3). • Remove screws (6) and rear cover (7). • Remove screws (4) and front covers (5). MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 2 of 8 Siemens AG Rev. 03 07.
  • Page 9: Removing The Swivel Arm Cover

    X808, connected to control panel (10)inside the shaft cover. CAUTION Be careful not to damage connectors X807 and X808 when remov- ing the covers. Siemens AG Register 5 RXB7-230.091.05 Page 3 of 8 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
  • Page 10: Removing The Compression Unit

    Fig. 4 Position of the protective conductors and the cable guide • Remove cable guide (1). • Remove earth screw (2) and ground cables (3). MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 4 of 8 Siemens AG Rev. 03 07.
  • Page 11 Pull out connectors X873 and X874, which are located behind the compression unit. • Loosen and remove screw 3, and remove the compression unit. Siemens AG Register 5 RXB7-230.091.05 Page 5 of 8 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
  • Page 12: Removing Other Components

    Remove the connection wires for the fan. • Remove the 4 screws (1/Fig. 6) and remove the fan unit (2/Fig. 6). Fig. 6 Removal of fan unit MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 6 of 8 Siemens AG Rev.
  • Page 13: Mammomat 3000 / Mammomat 3000 Ce Only

    Remove the table support cover (5/Fig. 8). • Loosen bellow (7/Fig. 8) by removing plastic guide blocks inside the bellow. Siemens AG Register 5 RXB7-230.091.05 Page 7 of 8 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
  • Page 14 2 - 8 Preliminary work This page intentionally left blank. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 8 of 8 Siemens AG Rev. 03 07. 98 TD SD 24 Medical Engineering...
  • Page 15: Remove The Connectors At The Mammomat 300/300 Ce Swivel Arm

    3 - 1 Remove the connectors at the MAMMOMAT 300/300 CE swivel X897 X855 X856 X860 X896 X870 X895 X894 4 emergency stop button cables (backside) RIGHT SIDE LEFT SIDE Fig. 1 • Remove all connectors in accordance with Fig. 1 and the list which follows (including the protective conductors).
  • Page 16: Remove The Connectors At The Mammomat 3000/3000 Ce Swivel Arm

    • Eject wire #2 from connector X860. • Remove the four wires from the STOP switch. • Remove the wire with connector X893. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 2 of 16 Siemens AG Rev. 03 07. 98...
  • Page 17 5 protective conductors swivel arm frame 2 fan connection wires on the fan *1 MAMMOMAT 3000/ 3000 CE only Siemens AG Register 5 RXB7-230.091.05 Page 3 of 16 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
  • Page 18: All Mammomats: Removing The High Voltage Cable

    Lift up the X-ray tube and remove the high voltage cable (1) by unscrewing the sleeve nut (2). • Secure the X-ray tube again with the mounting screws. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 4 of 16 Siemens AG Rev.
  • Page 19: Mammomat 300/300 Ce Only - Removing The Corrugated Hose

    Remove the component parts of the high voltage cable plug. • Cut the cable ties from the cable harnesses being removed. • Remove the cable from the swivel arm. MAMMOMAT 300/300 CE only - Removing the corrugated hose Fig. 5 R803 Fig. 6 •...
  • Page 20: Mammomat 3000/ 3000Ce Only - Removing The Corrugated Hose

    - The braided grounding wire should remain attached to the swivel arm. - Unfold the shielding and remove the semicircular cable clamp. • Pull the lead-through ring off the cable harness. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 6 of 16 Siemens AG Rev.
  • Page 21: All Mammomats: Additional Disassembly Work

    Remove only X893 (WING 1) Fig. 8 • Pull all cables out of the swivel arm as shown in Fig. 8. Siemens AG Register 5 RXB7-230.091.05 Page 7 of 16 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
  • Page 22: All Mammomats: Removing The Cable Harness From The Left Side Of The Stand

    Removing de cable harness All Mammomats: Removing the cable harness from the left side of the stand NOTICE for MAMMOMAT 300/ MAMMOMAT 300 CE: Some of the connectors below are omitted All connectors on D801, except flat cable to D802...
  • Page 23: On The Left Side Of The Stand Remove

    *1 omit for MAMMOMAT 3000 CE. MAMMOMAT 3000 CE only • Remove connector X802 in the base of the stand (transformer T802). Siemens AG Register 5 RXB7-230.091.05 Page 9 of 16 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
  • Page 24: All Mammomats: Disconnecting The Cable Harness At The Back Wall Of The Stand

    All Mammomats: Disconnecting the cable harness at the back wall of the stand NOTICE For MAMMOMAT 300/ MAMMOMAT 300 CE: Some of the connectors below are omitted. X887 Lift motor (Note the color before disconnecting X887) X888 and X826 Swivel motor...
  • Page 25: Removing The Connectors On The Left Side Of The Stand

    - Remove the 3 screws show in (2/Fig. 11) and remove the swivel motor (3/Fig. 11). Fig. 11 Siemens AG Register 5 RXB7-230.091.05 Page 11 of 16 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
  • Page 26: Mammomat 300 Ce And Mammomat 3000 Ce Only

    3 - 12 Removing de cable harness MAMMOMAT 300 CE and MAMMOMAT 3000 CE only Remove the screws from the connection device (1/Fig. 12) on the base of the stand and disconnect the following connectors: • X885 connection cable for the stereo evaluation unit.
  • Page 27: Removing The Cable Harness From The Stand

    Fig. 13 Fig. 14 • Loosen the hose clamp (1/Fig.13) for the corrugated hose on the lifting carriage.f for MAMMOMAT 300 CE and MAMMOMAT 3000 CE NOTICE The braided grounding wire remains attached to the lifting carriage. • Remove the angle bracket (1/Fig.14) for the corrugated hose.
  • Page 28: Removing The Cable Harness At The Generator

    X1 Serial interface. • X10, X11 Iontomat cable. • Protective conductor. Pull back all cables that lead from the generator to the base of the stand. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 14 of 16 Siemens AG Rev. 03 07.
  • Page 29: Mammomat 300 / Mammomat 3000 Only

    Removing de cable harness 3 - 15 MAMMOMAT 300 / MAMMOMAT 3000 only Fig. 16 • Remove the cable bracket (1/Fig.16) from the base of the stand. All Mammomats • Withdraw the generator cable into the stand. • Remove the cable harness.
  • Page 30: Rerouting The Wires

    (for harness 62 40 175). • The conductors at X885 are connected as follows: white brown green Fig. 17 MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 16 of 16 Siemens AG Rev. 03 07. 98...
  • Page 31: Routing The Cable Harness

    Route the cable (2/Fig. 2) in the lower part of the stand and secure it with cable ties. Fig. 2 Siemens AG Register 5 RXB7-230.091.05 Page 1 of 14 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
  • Page 32 • Route the branch with the connector for D806 on the lifting carriage as shown in 1/Fig. 3 and secure it with cable ties. Fig. 3 MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 2 of 14 Siemens AG Rev.
  • Page 33: Connecting The Cable Harness To The Left Side Of The Stand

    Connect X886 (inside of stand) X801 *1 not for MAMMOMAT 3000 CE Fig. 4 NOTICE Some connections may not be present on MAMMOMAT 300 / MAMMOMAT 300 CE. On MAMMOMAT 3000 CE: Plug X802 into the receptacle in the stand. Siemens AG Register 5 RXB7-230.091.05...
  • Page 34: Connecting The Cable Harness To The Back Wall Of The Stand

    X888 and X826 Swivel motor S882 X861 on D806 S881 Protective conductor X898 Connector for magnetic brakes S883 S884 Fig. 5 MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 4 of 14 Siemens AG Rev. 03 07. 98 TD SD 24 Medical Engineering...
  • Page 35: If The Swivel Motor Has Been Removed

    • Attach the cable bracket to the base of the stand (1/Fig. 6). Fig. 6 MAMMOMAT 300 CE / MAMMOMAT 3000CE only Connect the wires on the multi-purpose connection device in the stand: • Attach plug X885 using the screws and connect the wires to the stereo evaluation unit (if used).
  • Page 36 Installing the cable harness • Secure the angle bracket (2) to the corrugated hose with the screws. MAMMOMAT 300 CE / MAMMOMAT 3000 CE only Make sure that the shielding is folded over the lead-through ring NOTICE and that the semicircular cable clamp is in position. Also make sure that the braided grounding wire is between the hose clamp and the shielding.
  • Page 37: Mammomat 300/ Ce

    X856 X860 X896 X870 X895 X894 4 emergency stop button cables (backside) RIGHT SIDE LEFT SIDE Fig. 9 Siemens AG Register 5 RXB7-230.091.05 Page 7 of 14 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
  • Page 38: Mammomat 3000/ Ce

    Also make sure that the braided grounding wire is between the hose clamp and the shielding. • Install the stereo switch with the screws. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 8 of 14 Siemens AG Rev.
  • Page 39: Mammomat 300/ Ce

    Divide the cable harness and insert it. Route the cable harness so that it is free of coils and kinks. • Secure the cable harness with cable ties. Siemens AG Register 5 RXB7-230.091.05 Page 9 of 14 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
  • Page 40: Mammomat 3000/ Ce

    Secure the cable with cable ties. • Reinsert wire 2 into connector X860. • Insert all cables as shown in the Fig. and connect as many plugs as possible at this point. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 10 of 14 Siemens AG Rev.
  • Page 41: Installing X891 And X893

    Connecting the STOP switch STOP switch connections Connection no. white (WH) blue (BU) black (BK) yellow (YE) Fig. 15 Siemens AG Register 5 RXB7-230.091.05 Page 11 of 14 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
  • Page 42: Routing The Cable In The Swivel Arm

    (light barriers for measuring the +/- 10° tilt of the tube.) • Reinstall the rubber bellows over WING 1 / WING 2 (refer to Pos. 7 in Fig. 2 - 8) with the screws. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 12 of 14 Siemens AG Rev.
  • Page 43: Correct Positioning Of The Compression Unit

    Check the system functions according to the adjustment instructions (e.g. kV measurement, Iontomat test on both wings if fitted etc.). • Attach the generator cover. Siemens AG Register 5 RXB7-230.091.05 Page 13 of 14 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
  • Page 44 4 - 14 Installing the cable harness This page intentionally left blank. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 14 of 14 Siemens AG Rev. 03 07. 98 TD SD 24 Medical Engineering...
  • Page 45 Text changed 4-13 Text changed TD HSC 24 / Guggenmos TD SD 24 / Gross SMS Iselin / O’Donnell Siemens AG Register 5 RXB7-230.091.05 Page 1 of 2 MAMMOMAT 300 / 3000 Medical Engineering Rev. 03 07. 98 TD SD 24...
  • Page 46 5 - 2 Changes to previous version This page intentionally left blank. MAMMOMAT 300 / 3000 Register 5 RXB7-230.091.05 Page 2 of 2 Siemens AG Rev. 03 07. 98 TD SD 24 Medical Engineering...