Technical Data / Information
Ambient temperature: 10°C to 40°C resp. 50°F to 104°F (operation temp. of instrument)
Barometric pressure:
Humidity:
Humidity protection:
Over-temperature protection:
Dimensions:
Weight:
7.2
Transmitter
Frequency range:
Control of acoustic output:
Focusing:
Processing-channels:
Sound field parameters:
7.3
Receiver
Frequency range:
Focusing (with annular and multi-element):
Processing-channels:
Receive-apodization:
TGC:
Dynamic range:
7-4
-10°C to 40°C resp. 14°F to 104°F (stock and transport temp.)
working temp. of probes see probe user's manual
700 to 1060 hPa (operation condition)
700 to 1060 hPa (stock and transport condition)
30 to. 80% RH no condensation (operation condition)
0 to. 90% RH no condensation (stock and transport condition)
covered, no humidity protection
yes, provided by three fans
680 x 1000 x 1450 mm [width x depth x height]
26.7 x 39.4 x 56.1 in.
basic unit (without accessories) approx. 136 kg
Broad band system 1 to 16 MHz, automated adaptation to the probe
used.
Range: 32 dB, adjustable in 1dB steps
selectable transmit focal length and focal depth
512 channels
the declaration of the sound field parameters acc. to IEC1157 can be
furnished by the manufacturer on request. Sound field limiting
parameters on request.
Broad band system 1 to 16 MHz, automated adaptation to the probe used.
Sub-pixel based digital dynamic focusing system:
Accuracy of focus: +/- 3ns
Sample-rate: 60.0 MHz
High resolution Mode: 512 channels
yes
manual, 100 dB control range by gain knob and slider-pots
150dB
®
Voluson
730 - Instruction Manual
105838 Rev. 3