ABB ZDT Installation And Operating Manual page 23

High temperature version, for use with zgp2 probes
Table of Contents

Advertisement

...7.7 Set Up Outputs Page
Continued from previous page
00. 0 0
Retrans Zero
25 . 0 0
Retrans Span
-----
Hold Outputs NO
100. 0
Test Retrans (%)
00000
Alter Sec. Code
00000
Alter Cal. Code
ELECTRICAL CAL.
Retransmission Zero
Set the required retransmission zero value, in % oxygen units.
[0% to 20.00% (linear output)]
or
[0.01% to 0.25% (logarithmic output)].
Retransmission Span
Linear Output – Set the required retransmission span value, in % oxygen units.
[5% to 25.00%]
Logarithmic Output – The retransmission span is preset to two decades above
the zero value and cannot be adjusted. For example, if the zero is set to 0.2%,
the span is preset to 20.00% .
Hold Outputs
The retransmission and alarm outputs can be held to prevent inadvertent
operation during a test gas calibration sequence.
YES
Test Retransmission
The instrument transmits a test signal of 0, 25, 50, 75 and 100% of the
retransmission range selected above. The % test signal selected is shown on
the upper display.
Example: When the Retransmission Current Range is set to 0-20 (mA) and
the Test Retransmission signal is set to 50%, the retransmission output value
is forced to 10mA.
Alter Security Code
[00000 to 199999]
Set the security code used to gain access to secure parameters – see
Section 7.4.
Alter Calibration Code
[00000 and 199999]
Set the code used to gain access to the oxygen calibration facility – see
Sections 7.1 to 7.3.
Press
Press
to return to the top of the
Set Up Outputs Page.
to advance to the
Electrical Calibration Page.
7 PROGRAMMING
21

Advertisement

Table of Contents
loading

Table of Contents