Page 1
SYSTEM B FS ISORAD UCKY IAGNOST LIST OF DOCUMENTATION IN THIS BINDER: ⊗ SUBSYSTEM MANUAL BuckyDiagnost FS IsoRAD UNIT MANUAL manual/automatic collimator Note: ⊗indicates document present 4512 984 28681 REV AA 2004 Philips Medical Systems ALL RIGHTS RESERVED BD_FS_IsoRAD_01...
Page 2
INTRODUCTION AND TECHNICAL DATA Philips Medical Systems SERVICE MANUAL INSTALLATION SUBSYSTEM BuckyDiagnost FS IsoRAD FAULT FINDING 9890 010 83921 REPLACEMENT PROGRAMMINGS ADJUSTMENTS DMC Hamburg DRAWINGS Printed in Hamburg, Germany 2004 Philips Medical Systems ALL RIGHTS RESERVED SCHEMATIC DIAGRAMS 4512 984 28681 REV AA...
Page 3
BuckyDiagnost FS IsoRAD SERVICE MANUAL - - SUBSYSTEM BuckyDiagnost FS IsoRAD Author: M. Bierstedt Type No : 9890 010 83921 In case there are any questions concerning this manual, please send this LOPAD via fax to 49/(0)40/5078 2481 File: BD_FS IsoRAD_28681AA...
10 peak Shock pulse duration in msec Air pressure in Hecto-pascal 700 / 1100 700 / 1100 Acoustic noise level Air cooling IEC 950 1--2 (04.0) BuckyDiagnost FS IsoRAD 2004 Philips Medical Systems 1_Introduction and technical data_BW ALL RIGHTS RESERVED...
+10 ... 15% 1.4.4. Tools / Material required The FS IsoRAD is already installed and aligned in the shelter. For replacement activities see chapter 4. BuckyDiagnost FS IsoRAD (04.0) 1--3 2004 Philips Medical Systems 1_Introduction and technical data_BW ALL RIGHTS RESERVED...
Start of installation, operation and maintenance work and especially electrical work must only be executed by trained and authorized persons. This equipment must only be serviced by properly educated service specialists who have received general and system--specific training as performed by Philips Medical Systems. WARNING The system/component must be switched OFF during replacement work.
Tool Code Table Fixed Table Height adjustable Vertical Vertical Radiography Vertical Tiltable 1.9. Manual history Date Version Name Reason of changes 16.08.04 M. Bierstedt BuckyDiagnost FS IsoRAD (04.0) 1--5 2004 Philips Medical Systems 1_Introduction and technical data_BW ALL RIGHTS RESERVED...
Page 10
BuckyDiagnost FS IsoRAD INSTALLATION Installation The FS IsoRAD is already installed and aligned in the shelter. For replacement activities refer to section 4. BuckyDiagnost FS IsoRAD (04.0) 2--1 2004 Philips Medical Systems 2_Installation_04_BW ALL RIGHTS RESERVED...
Page 11
BuckyDiagnost FS IsoRAD FAULTFINDING Fault finding Use manual SMCM BuckyDiagnost TH2, manual Code No. 4512--984--2565x. BuckyDiagnost FS IsoRAD (04.0) 3--1 2004 Philips Medical Systems 3_Fault_Finding_BW ALL RIGHTS RESERVED...
Page 12
......4--10 BuckyDiagnost FS IsoRAD (04.0) 4--0.1 2004 Philips Medical Systems Repl_4_inh ALL RIGHTS RESERVED...
Page 13
4 = Cable support cover 5 = Electronic rack cover 6 = Column base rear cover 7 = Column base side cover, rear 8 = Column top cover BuckyDiagnost FS IsoRAD (04.0) 4--1 2004 Philips Medical Systems 4_Replacements_mit_12NC_04 ALL RIGHTS RESERVED...
Page 14
S Install the rail anchor (1). S Ensure that the wheel runs free. -- If not, change the position of the distance plates. S Re-install the cover. 4--2 (04.0) BuckyDiagnost FS IsoRAD 4_Replacements_mit_12NC_04 2004 Philips Medical Systems ALL RIGHTS RESERVED...
Page 15
S Install the brake according to chapter 1.2.1 “Replacement of the longitudinal brake”. S Install the anchor rail. S Re-install the covers. S Check the functions. BuckyDiagnost FS IsoRAD (04.0) 4--3 2004 Philips Medical Systems 4_Replacements_mit_12NC_04 ALL RIGHTS RESERVED...
Page 16
S Loosen the four screws (1) and replace the base plate installed micro switch. -- Ensure the correct function of the micro switch. S Check the function of the system. S Re-install the cover. 4--4 (04.0) BuckyDiagnost FS IsoRAD 4_Replacements_mit_12NC_04 2004 Philips Medical Systems ALL RIGHTS RESERVED...
Page 17
S Remove the base plate installed micro switch according to chapter 1.3.2. S Remove the screws (1) and replace the catch. S Install the base plate installed micro switch according to chapter 1.3.2. BuckyDiagnost FS IsoRAD (04.0) 4--5 2004 Philips Medical Systems 4_Replacements_mit_12NC_04 ALL RIGHTS RESERVED...
Page 18
Code No. 4512 13 1441. Replacement of the alpha switch plate S Switch the system OFF. S Turn the control handle clockwise to its end position. S Remove the cover (1). 4--6 (04.0) BuckyDiagnost FS IsoRAD 4_Replacements_mit_12NC_04 2004 Philips Medical Systems ALL RIGHTS RESERVED...
Page 19
S Remove the nut (1). S Remove the upper screw of the alpha switch plate (1) S Pull out the alpha switch plate and remove the wires. BuckyDiagnost FS IsoRAD (04.0) 4--7 2004 Philips Medical Systems 4_Replacements_mit_12NC_04 ALL RIGHTS RESERVED...
Page 20
1.3.6. Replacement of the transverse brake Code no. 4512- -130- -7291. Replacement of the transverse brake S Switch the system OFF. S Remove the bottom cover of the tube carrier arm. 4--8 (04.0) BuckyDiagnost FS IsoRAD 4_Replacements_mit_12NC_04 2004 Philips Medical Systems ALL RIGHTS RESERVED...
Page 21
S Solder the cable. S Fix the cable at the tube carrier arm. S Fix the end stopper. S Check the function of the brake. S Re-install the cover. BuckyDiagnost FS IsoRAD (04.0) 4--9 2004 Philips Medical Systems 4_Replacements_mit_12NC_04 ALL RIGHTS RESERVED...
Page 22
PCB on the control handle Note After replacement of the CU2 proceed height adjustment. Refer to chapter 10.2 ’Adjustment of the ceiling suspension height’ in manual X- -Scope. 4--10 (04.0) BuckyDiagnost FS IsoRAD 4_Replacements_mit_12NC_04 2004 Philips Medical Systems ALL RIGHTS RESERVED...
Page 23
Programming of the UZ field 4512 108 0738x: Coding field S Open link W1 (1). +24V UZX7:5 UZX2:2 +24V 1.2. Software programmings with X-Scope Refer to manual X-Scope in SRM BuckyDiagnost FS IsoRAD. BuckyDiagnost FS IsoRAD (04.0) 5--1 2004 Philips Medical Systems 5_Programmings ALL RIGHTS RESERVED...
Page 24
............6--3 BuckyDiagnost FS IsoRAD (04.0) 6--0.1 2004 Philips Medical Systems Adjstment 6_inh ALL RIGHTS RESERVED...
Page 25
S Align the tube by loosen screws at the holder or of the tube. Respect the direction of vary. -- Fasten screws, then repeat procedure. S Remove the mirror. BuckyDiagnost FS IsoRAD (04.0) 6--1 2004 Philips Medical Systems 6_Adjustments ALL RIGHTS RESERVED...
Page 26
S Tighten screws, then repeat procedure. Fine Adjustment: S Loosen the two fixing screws (4) of the center laser. S Adjust the laser. S Tighten the two fixing screws (4). 6--2 (04.0) BuckyDiagnost FS IsoRAD 6_Adjustments 2004 Philips Medical Systems ALL RIGHTS RESERVED...
Page 27
S Install the removed control handle grip on the rear side Note Pay attention to the laser wires, that they are not damaged by attaching the grip! If necessary, fix the wires with adhesive tape. BuckyDiagnost FS IsoRAD (04.0) 6--3 2004 Philips Medical Systems 6_Adjustments ALL RIGHTS RESERVED...
Page 28
BuckyDiagnost FS IsoRAD mechanical dimensions Z--1.1 BuckyDiagnost FS IsoRAD floor rails mechanical dimensions Z--1.2 Installation of floor rail and wall rail Z--1.3 Cable outlet Z--1.7 Labelling 2Z--10 BuckyDiagnost FS IsoRAD (04.0) Z--0.1 2004 Philips Medical Systems drawings_Z- -0_1 ALL RIGHTS RESERVED...
Page 29
1000 1212 1000 BuckyDiagnost FS IsoRAD Mechanical dimensions BuckyDiagnost FS ISO RAD (04.0) Z-1.1 C Philips Medical Systems...
Page 30
4900 BuckyDiagnost FS Iso RAD Floor rail 4900mm Mechanical dimensions BuckyDiagnost FS ISO RAD (04.0) Z-1.2 C Philips Medical Systems...
Page 31
BuckyDiagnost FS Iso RAD Installation of floor rail and wall rail Z-1.3 (04.0) BuckyDiagnost FS ISO RAD Philips Medical Systems...
Page 33
System labels Tube holder assy Labeling as indicated UL / CSA in survey of tubes Logo Address Typen No. Electr. data IEC 60601-1 IEC 60601-2-32 Identifications label Stand FS BuckyDiagnost FS Labelling 4512 983 07571 (b/04.0) 2Z-10 C Philips Medical Systems...
Page 34
BuckyDiagnost FS IsoRAD Earthing diagram Z2--1 BuckyDiagnost FS IsoRAD Wiring diagram Z2--2 BuckyDiagnost FS CAN bus cable Z2--3 BuckyDiagnost family control handle with display and SID laser Z2--2.2 BuckyDiagnost FS IsoRAD (04.0) Z--0.2 2004 Philips Medical Systems Schematic_Wiring_Z- -0_2 ALL RIGHTS RESERVED...
Page 35
= Position switch for lateral movement (x-direction) = Tube housing support = Switch for alpha rotation position = PCB connection board = Tube UQBS2 UQBS1 BuckyDiagnost ISO RAD Survey of units BuckyDiagnost FS ISO RAD (04.0) Z1-1 C Philips Medical Systems...
Page 36
X5:2 X14:2/5 X4:3 X4:4 24V+ X13:1 X1:3 yw/gn 3.2A slow green X13:2 X12:1 LA Collimator +24V/12V UBN3 Tomo 220V X12:3 X13:3 12V+ X13:4 BuckyDiagnost FS Power supply UAN1 BuckyDiagnost FS (03.0) Z1-2 C 2001 Philips Medical Systems ALL RIGHTS RESERVED...
Page 37
Telescope arm Telescope arm Movement Switch movement Catch Y and SID Column stand Y-Brake ~230V Potenti ometer Y-Brake Column 230V Mains filter BuckyDiagnost FS ISO RAD with sensing System cabling BuckyDiagnost FS ISO RAD (04.0) Z1-3 C Philips Medical Systems...
Page 38
X1:07 X11:07 magnet CLUTCH_ALPHA X4:3 driver tube driver BuckyDiagnost FS X5:5 diaphragm TUBE_SEL1 X5:3 Detailed schematic PO 26V remote light driver X5:6 Generator driver TUBE_SEL2 X5:4 BuckyDiagnost FS (04.0) Control handle Z1-4 C 1994 Philips Medical Systems ALL RIGHTS RESERVED...
Page 39
CH_CLK GR_CLK driver driver CH_DATA GR_DATA 4512 108 1036x GD_LS Control grip CS Nicol 4512 201 0234x Control grip CS Nicol-SID 4512 201 0233x (with option SID laser) BuckyDiagnost FS (03.0) Z1-5 C 1994 Philips Medizin Systeme ALL RIGHTS RESERVED...
Page 40
Coding field +24V UZX7:5 UZX2:2 +24V Coding field 4512 108 0738. BuckyDiagnost FS (04.0) Z1-7 C Philips Medical Systems...
Page 41
Screw Collimator UAN1 X1:3 Tube Main Filter Tube assembly Power Rack Screw H.T. generator Screw Corrugated hose Corrugated hose support Screw Column stand Catch BuckyDiagnost FS ISO RAD Earthing diagram (04.0) BuckyDiagnost FS ISO RAD Z2-1 C Philips Medical Systems...
Page 42
Position Z UEX4:7 USBX101 12Vdc 24Vdc UACY1 UEX1 UEX4:5 Control handle Brake Z-Direction UEX4:6 BuckyDiagnost FS ISO RAD UZX14 UZX14 Rotation Beta switch SZ2X13 Wiring diagram Z2-2 (04.0) BuckyDiagnost FS ISO RAD C 1997 Philips Medical Systems DMC ALL RIGHTS RESERVED...