Resetting the Totalizers

Resetting Totalizers Using
the Internal Keypad
Resetting Totalizers Using
an External Switch
Displaying Data
To reset the totalizers, you can use either the internal keypad or an
external switch. When resetting the totalizers, the totals for both
channels are reset. Use the appropriate section below to reset the
totalizers.
See Figure A-4 on page A-4 in Appendix A, Menu Maps. Then,
1. Press [ESC] , [ENTER] , [ESC] in sequence to enter the user program.
2. Press [ ] until RESET appears and press [ENTER] .
3. Press [ ] to select YES or NO and press [ENTER] .
4. To leave the User Program, press [ ESC ].
If you select YES , the AT868 resets the totalizers for both channels. If
you select NO , the totalizers continue to count.
While the meter is operating, press and hold the external reset switch
for 1 second. Both totalizers will reset to zero and will then
immediately begin to resume counting.
IMPORTANT: The external totalizer reset switch must be configured
as described in Chapter 2, Programming Site Data.