Philips AVENT SCH740 Manual page 8

Hide thumbs Also See for AVENT SCH740:
Table of Contents

Advertisement

Available languages

Available languages

Electro
static
dis
charge
(ESD)
IEC
61000-4
-2
Power
frequen
cy
(50/60H
z)
magnet
ic field
IEC
61000-4
-8
Table 4 Guidance and manufacturer's
declaration – electromagnetic immunity
–for ME equipment and ME systems that
are not life supporting
Guidance and manufacturer's declaration –
electromagnetic immunity .The device is intended
for use in the electromagnetic environment
specified below. The customer or the user of the
device should assure that it is used in such an
environment.
IMMUNITY
test
Conducted RF
IEC
61000-4-6
Radiated RF
IEC 61000-4-3
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any part of
the device, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance:
d = 1.2 ÖP
d = 1.2 ÖP 80 MHz to 800MHz
d = 2.3 ÖP 800 MHz to 2.5 GHz
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey (a),
should be less than the compliance level in each
frequency range (b).
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all
situations. Electromagnetic propagation is affected
by absorption and reflection from structures,
objects and people.
(a) Field strengths from fixed transmitters, such as
base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field
strength in the location in which the device is used
exceeds the applicable RF compliance level above,
the device should be observed to verify normal
operation. If abnormal performance is observed,
additional measures may be necessary, such as re-
orienting or relocating the device.
(b) Over the frequency range 150 kHz to 80 MHz,
field strengths should be less than 3V/m.
Table 6 Recommended separation
distances between portable and mobile
RF communications equipment and the
ME equipment or ME system – for ME
equipment and ME systems that are not
life supporting
Recommended separation distances between
portable and mobile RF communications
equipment and the device.
The device is intended for use in an
electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the
user of the device can help prevent
electromagnetic interference by maintaining a
minimum distance between portable and mobile
RF communications equipment (transmitters) and
the device as recommended below, according to
the maximum output power of the
communications equipment.
 
Rated
maximum
output
power of
transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power
not listed above, the recommended separation
distance d in metres (m) can be estimated using
the equation applicable to the frequency of the
transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to
the transmitter manufacturer.
±6 kV
contact
±8 kV air
3A/m
IEC 60601 TEST
LEVEL
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to 2.5
GHz
Separation distance according to
frequency of transmitter (m)
150 kHz to
80 MHz
d = 1.2 Ö P
0.12
0.37
1.2
3.7
12
±6 kV
Floors should be
contact
wood, concrete or
±8 kV
ceramic tile. If
air
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%.
3A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in
a typical
commercial or
hospital
environment.
80 MHz
to 800
MHz
d = 1.2 Ö P
0.12
0.37
1.2
3.7
12
Compliance
level
3 Vrms
3 V/m
800 MHz
to 2.5 GHz
d = 2.3
0.2
0.74
2.3
7.4
23