Problem
The security lever doesn't move to the far
left position.
Revision B
Troubleshooting: General
Cause
Cables or other objects may be blocking
the monitor's AC power or communication
ports.
The monitor's expansion pcb may be
damaged.
The monitor may not be positioned
correctly on the docking station.
Foreign objects may be blocking the
monitor's expansion pcb.
The AC power/communication mechanism
has failed.
Dash Port Docking Station
2000966-037
Solution
Remove the monitor from the docking
station and remove any cables or other
objects from the connector ports located on
the bottom of the monitor.
Replace the monitor's damaged expansion
pcb. See the appropriate Dash Patient
Monitor's Service Manual.
Remove and reposition the monitor.
Remove foreign objects blocking the
expansion PCB.
Remove the monitor from the docking
station. Manually depress the locking lever
pin and slide the security lever to the far left
position.
The AC power/communication mechanism
should extend out of the docking station's
base and lock into position. If it doesn't,
complete the following tasks.
1. Verify the plunger meets the force
specifications identified on page 2 of the
"Exploded View PN 2004740-001,
Revision D"
2. Complete,
"Replacing the Connector
Assembly or the Bottom Housing"
page 5-5.