Asus LU700 Troubleshooting Manual page 5

Ultrasound imaging system
Hide thumbs Also See for LU700:

Advertisement

2.
To be ensured that all the coupling gel and other visible substances from the probe is removed by
wiping with a clean paper towel. If necessary, to remove material dried to the surface, the cloth can
be moistened with lukewarm water.
3.
It shall inspect the probe's lens and casing after each use. To check out any damage that would allow
liquid to enter the probe. If the user found a probe damage, the probe shall not be placed into any
liquid (e.g., for disinfection) and shall not be used until it has been inspected and repaired/replaced
by ASUS.
Recommendations for disinfecting the ultrasound probe (After cleaning):
1.
Spray 70% Isopropyl Alcohol onto the surface of probe head.
2.
Repeat step one for two or three times.
3.
Wipe out the disinfectant with a clean paper towel.
No image or abnormal display after connected
- When LU700 is connected normally but without ultrasonic image on screen or abnormal lines on the
image, please check if there's electromagnetic interference around and restart the LU700.
Freeze button abnormality
- When clicking the freeze button cannot enter the pause state under scan mode, please restart the
LU700 and the App.
- If you still can't enter the pause state, please contact ASUS Customer Service Center.
Appearance defect
- Please contact ASUS Customer Service Center if there are defects in the appearance of the new product,
such as collision scratches or peeling off of the rubber.
- If it is artificial damage, it is not within the scope of warranty, users could contact ASUS Customer
Service Center to pay for maintenance.