DIRECTION 5854713-100, REVISION 2
10-5-2

Functional Checks

NOTE:
Refer also to Chapter 4 - General Procedures and Functional Checks, for additional details about the
functional checks described in this section.
The functional checks take approximately 60 minutes to perform. Refer to the Venue Fit™ User Manual
whenever necessary.
10-5-2-1

System Checks

Step
1
B-Mode
2
CF-Mode
3
Doppler Modes
4
M-Mode
5
Probe Elements
6
Applicable Software Options
8
Monitor
9
Peripherals
Table 10-5
System Functional Checks
Item (or Mode)
Verify basic B-Mode (2D) operation.
Check the basic Ultrasound system controls that affect this mode of operation.
Verify basic CF-Mode (Color Flow Mode) operation.
Check the basic Ultrasound system controls that affect this mode of operation.
Verify basic Doppler operation (PW and CW if available).
Check the basic Ultrasound system controls that affect this mode of operation.
Verify basic M-Mode operation.
Check the basic Ultrasound system controls that affect this mode of operation.
Perform an Element Test on each probe to verify that all the probe elements and system
channels are functional.
Verify the basic operation of all optional modes such as Contrast. Check the basic
Ultrasound system controls that affect each option's operation.
Verify basic monitor display functions.
See:
Chapter 10 - Care and Maintenance
Peripheral/Option Checks
on page 10 - 8.