Guidance and manufacturer's declaration – electromagnetic immunity
The FPI is intended for use in the electromagnetic specified below.
The customer or the user of the FPI should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC61000-4-2
Electrical fast
transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC61000-4-11
Power frequency
(50/60Hz)
magnetic field
IEC61000-4-8
Note:
(1) U
is the a.c.mains voltage prior to application of the test level.
T
(2) This FPI can only allow to use the designated cables and components.
If the different cables or components are used, they may deteriorate the performance of
recommended separation distance between portable and mobile RF communications
equipment and this system.
IEC60601
test level
±6kV contact
±2/4kV contact
±8kV air
±2/4kV air
±2 kV for power
supply lines
±1kV for input/
output lines
±1kV differential
mode
±0.5kV differential
mode
±2kV common
mode
<5 % U
T
(>95 % dip in, U
)
T
for 5 sec
3A/m
Compliance level
±6kV contact
±2/4kV contact
±8kV air
±2/4 kV air
±2kV for power
supply lines
±1kV for input/
output lines
±1kV differential
mode
±0.5kV differential
mode
±2kV common
mode
<5 % U
T
(>95 % dip in, U
)
T
for 5 sec
3A/m
-14-
Electromagnetic environment -
guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material.
The relative humidity should be at
least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical hospital environment.
Mains power quality should be that
of a typical hospital environment. If
the user of the FPI requires
continued operation during power
mains interruptions, it is
recommended that the FPI be
powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in atypical
commercial or hospital
environment.