4-6. How to attach/release the magnetic position sensor and the magnetic position sensor attachment

This section provides how to attach the magnetic position sensor and the magnetic position sensor attachment to the
probe.

4-6-1. How to attach the magnetic position sensor and the magnetic position sensor attachment

(1) Confirm that the magnetic position sensor attachment is sterilized or disinfected.
(2) Insert the magnetic position sensor into the magnetic position sensor attachment with the correct direction as
shown in Figure 8.
(3) Attach the magnetic position sensor attachment to the probe as shown in Figure 9.
Caution
Never attach the magnetic position sensor attachment to the probe in the incorrect direction,
otherwise it may result in false diagnosis.
Be careful not to put your fingers between the magnetic position sensor attachment and the
probe when attaching the attachment to the probe.
Do not use the magnetic position sensor attachment if it fell on the floor, otherwise there is
a risk of infection. Stop using it and perform the reprocessing procedure described in the
separate instruction manual "Cleaning, Disinfection and Sterilization" and make preparations
before use described in this manual.
Direction mark
of the sensor
Magnetic
Position Sensor
Magnetic Position
Sensor Attachment
Figure 8 How to attach Magnetic Position Sensor
Figure 9 How to attach Magnetic Position Sensor Attachment