Table 5-10 Alarm Adjustment Procedure - Honeywell 7866 Operation And Maintenance Manual

Triple range digital h2 and co2 analyzer/indicator and gas sampling panel
Table of Contents

Advertisement

Adjustment of the Range 3 Alarm Limits
A four-digit password may be required to adjust the Alarm limits. If the password is unknown, consult
your designated plant technician for the password. If you have forgotten the password, see Appendix B for
the Security Bypass Procedure.
Alarm Limits may be adjusted while the 7866 Digital Controller is operating. Care should be taken to
minimize the effects of entering out-of-range values.
Step
Press
1
SET UP
2
FUNCTION
3
or
FUNCTION
4
5
or
6
FUNCTION
7
or
FUNCTION
8
The 7866 should be supplemented with additional safety systems where possible and appropriate. Pressure loss
detectors and redundant gas loss detectors are recommended in critical applications. The user always assumes the
responsibility of the correctness of the safety system design. It is always recommended that the installer, designer,
and user consult with their insurance company, local authorities, and industry guidelines on such matters.
The alarm will activate when the measured percentage of hydrogen falls below the setpoint. A deadband, also called
"hysteresis", will prevent clearing of the alarm until the measured Hydrogen is above the setpoint by 0.5 percent.
Failure to comply with these instructions could result in death or serious injury.
9/06
7866 Triple Range Digital H

Table 5-10 Alarm Adjustment Procedure

until you see:
Upper Display:
SETUP
Lower Display:
ALARMS
You will see ALARM1LO in the Lower Display. This is the Alarm 1 setpoint, low
alarm type.
Enter limit value between 90.0 % to 100.0 %.
(Default = 98.0 %)
You will see ALARM2LO in the Lower Display. This is the Alarm 2 setpoint, low
alarm type
Enter limit value between 90.0 % to 100.0 %.
(Default = 90.0 %)
You will see HYSTERiS in the Lower Display. This is the Alarm Hysteresis which
sets the deactivation range for both alarms.
Enter hysteresis value between 0 % to 100.0 %.
(Default = 0.5 %)
Saves selections to nonvolatile memory.
and CO
Analyzer/Indicator - Operation and Maintenance
2
2
Action
Calibration
57